Determination of copper, manganese and nickel in biological samples and sea-water by flow injection on-line sorption preconcentration in a knotted reactor coupled with electrothermal atomic absorption spectrometry

被引:95
作者
Ivanova, E [1 ]
Benkhedda, K [1 ]
Adams, F [1 ]
机构
[1] Univ Instelling Antwerp, Dept Chem, B-2510 Wilrijk, Belgium
关键词
copper; manganese; nickel; flow injection on-line preconcentration; knotted reactor; electrothermal atomic absorption spectrometry; biological samples; sea-water;
D O I
10.1039/a709150h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection system with on-line sorption preconcentration of copper, nickel and manganese with ammonium pyrrolidinedithiocarbamate (APDC) or 8-hydroxyquinoline (HQ) in a knotted reactor (KR) coupled with electrothermal atomic absorption spectrometry was developed. APDC offered better performance characteristics than HQ for the preconcentration of copper and nickel but could not be employed for manganese. The performance characteristics of the determination of copper and nickel with APDC were: detection limits (3 sigma), 6.0 and 7.6 ng l(-1), precision (RSD) 2.5% for 0.2 mu g l(-1) and 1.5% for 0.2 mu g l(-1) and enhancement factors 44 and 21 using a 30 s preconcentration time. The corresponding parameters for the determination of manganese dth HQ were detection limit 29.0 ng l(-1), precision (RSD) 2.9% for 0.4 mu g(-1) manganese and enhancement factor 8 with a 30 s preconcentration time. The sampling frequency was 26 h(-1). The accuracy of the method was demonstrated by the analysis of certified reference materials and by recovery measurements on spiked samples.