Positive correlation between Tono-Pen intraocular pressure and central corneal thickness

被引:84
作者
Dohadwala, AA [1 ]
Munger, R [1 ]
Damji, KF [1 ]
机构
[1] Univ Ottawa, Inst Eye, Ottawa Gen Hosp, Ottawa, ON, Canada
关键词
D O I
10.1016/S0161-6420(98)91029-6
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To examine the relationship between intraocular pressure (IOP) readings taken by the Tone-Pen tonometer (Mentor O&O, Norwell, MA) and central corneal thickness (CCT). Design: Prospective cross-sectional population study. Participants: There were 651 eyes of 332 healthy subjects. Main Outcome Measures: A questionnaire was given to each subject requesting information on gender, age, race, and other factors that can influence IOP, The IOP then was measured using the Tone-Pen followed by measurements of CCT using an ultrasonic pachymeter, Results: The IOP was found to increase by 2.9 mmHg/100 mu m CCT in males and 1.2 mmHg/100 mu m in females. For males, CCT was found to be statistically significant in predicting IOP (P < 0.001 in the right and left eyes) and diabetes was of borderline significance (P = 0.012 in the right eye, P = 0.089 in the left eye). For females, CCT was of borderline significance (P = 0.064 in the right eye, P = 0.019 in the left eye). In females, a family history of glaucoma (P = 0.021 in the right eye, P = 0.022 in the left eye) and hypertension (P = 0.010 in the right eye, P = < 0.001 in the left eye) were also significant in the prediction of IOP, Race was found to be a significant predictor of CCT (P < 0.001 in both right and left eyes) for both males and females. Conclusion: Clinicians should be aware that, as with the Goldmann applanation tonometer, the Tone-Pen has a systematic error in IOP readings caused by its dependence on CCT, Tone-Pen IOP readings are positively correlated to CCT in males and, to a lesser extent, in females as well. The CCT measurements should be considered to ensure proper interpretation of IOP measurements in the diagnosis and management of disorders in which the CCT or IOP readings are outside normal limits.
引用
收藏
页码:1849 / 1854
页数:6
相关论文
共 29 条
[1]   Ocular hypertension and central corneal thickness [J].
Argus, WA .
OPHTHALMOLOGY, 1995, 102 (12) :1810-1812
[3]   THE TONO-PEN - A MANOMETRIC AND CLINICAL-STUDY [J].
BOOTHE, WA ;
LEE, DA ;
PANEK, WC ;
PETTIT, TH .
ARCHIVES OF OPHTHALMOLOGY, 1988, 106 (09) :1214-1217
[4]   THE DISTRIBUTION OF INTRAOCULAR PRESSURES IN THE GENERAL-POPULATION [J].
COLTON, T ;
EDERER, F .
SURVEY OF OPHTHALMOLOGY, 1980, 25 (03) :123-129
[5]  
DIELEMANS I, 1995, OPHTHALMOLOGY, V102, P54
[6]  
EDERER F, 1973, ARCH OPHTHALMOL-CHIC, V89, P1
[7]  
EHLERS N, 1976, ACTA OPHTHALMOL, V54, P774
[8]  
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34
[9]  
FRENKEL REP, 1988, ARCH OPHTHALMOL-CHIC, V106, P750
[10]  
HANSEN FK, 1971, ACTA OPHTHALMOL, V49, P82