Direct thrombolytic therapy for intraventricular thrombosis in patients with the Jarvik 2000 left ventricular assist device

被引:47
作者
Delgado, R
Frazier, OH
Myers, TJ
Gregoric, ID
Robertson, K
Shah, NA
Kar, B
机构
[1] Texas Heart Inst, Cardiopulm Transplant Serv, Houston, TX 77030 USA
[2] Texas Heart Inst, Cardiovasc Res Labs, Houston, TX 77030 USA
关键词
D O I
10.1016/j.healun.2003.10.023
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
One of the complications that can occur with continuous, axial-flow left ventricular assist devices (LVADs). is thrombosis within the left ventricle, adjacent to the device's inflow conduit, which may cause inflow obstruction and recurrent heart failure. We describe 2 cases in which we used a catheter. to continuously infuse recombinant tissue plasminogen activator (tPA) into the left ventricle until signs of successful thrombolysis was achieved. By monitoring the result and administering only as much tPA as necessary to achieve thrombolysis, we were able to successfully lyse the obstructing thrombus with a minimal dose of tPA without causing any significant bleeding problems. This technique may be useful for managing this potentially serious complication while minimizing the risk of treatment. Copyright (C) 2005 by the International Society for Heart and Lung Transplantation.
引用
收藏
页码:231 / 233
页数:3
相关论文
共 3 条
[1]   Initial clinical experience with the Jarvik 2000 implantable axial-flow left ventricular assist system [J].
Frazier, OH ;
Myers, TJ ;
Gregoric, ID ;
Khan, T ;
Delgado, R ;
Croitoru, M ;
Miller, K ;
Jarvik, R ;
Westaby, S .
CIRCULATION, 2002, 105 (24) :2855-2860
[2]   Clinical experience with the MicroMed DeBakey ventricular assist device [J].
Noon, GP ;
Morley, DL ;
Irwin, S ;
Abdelsayed, SV ;
Benkowski, RJ ;
Lynch, BE .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S133-S138
[3]  
ROSS AM, 1993, NEW ENGL J MED, V329, P1615