Homogeneity and diameter of linear lesions induced with multipolar ablation catheters: In vitro and in vivo comparison of pulsed versus continuous radiofrequency energy delivery

被引:10
作者
Erdogan, A [1 ]
Grumbrecht, S [1 ]
Carlsson, J [1 ]
Roederich, H [1 ]
Schulte, B [1 ]
Sperzel, J [1 ]
Berkowitsch, A [1 ]
Neuzner, J [1 ]
Pitschner, HF [1 ]
机构
[1] Max Planck Inst Physiol Res, Dept Cardiol Electrophysiol, Kerckhoff Clin, D-61231 Bad Nauheim, Germany
关键词
linear lesions; catheter ablation; pulsed energy delivery;
D O I
10.1023/A:1026538204579
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: For invasive treatment of atrial fibrillation, linear lesions induced with multipolar ablation catheters (MAC) are needed to prevent recurrence. The aim of the study was to compare the efficacy of pulsed versus continuous radiofrequency (RF)-energy delivery using MAC. Methods: In vitro tests were performed using endomyocardial preparations of fresh pig hearts in a 10-liter-bath of physiologic saline solution (37 degreesC) at constant flow conditions (1.5 l/min). The MAC were placed with a constant pressure of 20 ponds onto the endocardium. The energy (generator: Osypka HAT 200 S) was delivered either pulsed (4 electrodes simultaneously, 5 ms duty-cycle) or continuously (each electrode separately). In vivo experiments were performed in 6 anesthetized pigs using fluoroscopic positioning of MAC at 40 different intracardial positions and with similar conditions as in vitro experiments. Lesion volume (LV) was calculated after measuring lesion diameter with a microcaliper. The homogeneity of the lesions (LH) was classified from 1-4; with 1 as highest homogeneity. Results: Pulsed energy delivery produced more homogeneous linear lesions in significantly less time. There was no difference in electrode temperature values (50.2 +/- 0.8 and 51.3 +/- 1.4 degreesC) in vitro and in vivo. In the in vivo experiments, lesion depth and calculated lesion volume were less in both modes of energy delivery but pulsed energy delivery was superior regarding lesion depth and homogeneity. Conclusion: With pulsed energy delivery it is possible to create linear lesions of significantly greater homogeneity. Moreover, larger lesions are induced in less time by pulsed energy delivery in vitro and in vivo.
引用
收藏
页码:655 / 661
页数:7
相关论文
共 20 条
[1]  
Calkins H, 1999, AM J CARDIOL, V83, p227D
[2]   RADIOFREQUENCY ABLATION OF THE INFERIOR VENA-CAVA - TRICUSPID-VALVE ISTHMUS IN COMMON ATRIAL-FLUTTER [J].
COSIO, FG ;
LOPEZGIL, M ;
GOICOLEA, A ;
ARRIBAS, F ;
BARROSO, JL .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (08) :705-709
[3]   FROM FISHERMAN TO FIBRILLATION - AN UNBROKEN LINE OF PROGRESS [J].
COX, JL ;
BOINEAU, JP ;
SCHUESSLER, RB ;
KATER, KM ;
LAPPAS, DG .
ANNALS OF THORACIC SURGERY, 1994, 58 (04) :1269-1273
[4]   SUCCESSFUL SURGICAL-TREATMENT OF ATRIAL-FIBRILLATION - REVIEW AND CLINICAL UPDATE [J].
COX, JL ;
BOINEAU, JP ;
SCHUESSLER, RB ;
FERGUSON, TB ;
CAIN, ME ;
LINDSAY, BD ;
CORR, PB ;
KATER, KM ;
LAPPAS, DG .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1991, 266 (14) :1976-1980
[5]   Temperature monitoring during radiofrequency ablation [J].
Dinerman, JL ;
Berger, RD ;
Calkins, H .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (02) :163-173
[6]   THERMAL CATHETER DISRUPTION DURING CLOSED-CHEST RADIOFREQUENCY ABLATION OF THE ATRIOVENTRICULAR-CONDUCTION SYSTEM [J].
FROHNER, KJ ;
PODCZECK, A ;
HIEF, C ;
NURNBERG, M ;
STEINBACH, KK .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (06) :719-723
[7]   Interrelation of tissue temperature versus flow velocity in two different kinds of temperature-controlled catheter radiofrequency energy applications [J].
Gumbrecht, S ;
Neuzner, J ;
Pitschner, HF .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 1998, 2 (02) :211-219
[8]   OBSERVATIONS ON ELECTRODE-TISSUE INTERFACE TEMPERATURE AND EFFECT ON ELECTRICAL-IMPEDANCE DURING RADIOFREQUENCY ABLATION OF VENTRICULAR MYOCARDIUM [J].
HAINES, DE ;
VEROW, AF .
CIRCULATION, 1990, 82 (03) :1034-1038
[9]   Right and left atrial radiofrequency catheter therapy of paroxysmal atrial fibrillation [J].
Haissaguerre, M ;
Jais, P ;
Shah, DC ;
Gencel, L ;
Pradeau, V ;
Garrigues, S ;
Chouairi, S ;
Hocini, M ;
LeMetayer, P ;
Roudaut, R ;
Clementy, J .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (12) :1132-1144
[10]   COAGULATION OF VENTRICULAR MYOCARDIUM USING RADIOFREQUENCY ALTERNATING-CURRENT - BIO-PHYSICAL ASPECTS AND EXPERIMENTAL FINDINGS [J].
HAVERKAMP, W ;
HINDRICKS, G ;
GULKER, H ;
RISSEL, U ;
PFENNINGS, W ;
BORGGREFE, M ;
BREITHARDT, G .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (01) :187-195