Selective stimulation of cat sciatic nerve using an array of varying-length microelectrodes

被引:249
作者
Branner, A
Stein, RB
Normann, RA
机构
[1] Univ Utah, Dept Bioengn, Ctr Neural Interfaces, Salt Lake City, UT 84112 USA
[2] Univ Alberta, Div Neurosci, Edmonton, AB T6G 2S2, Canada
关键词
D O I
10.1152/jn.2001.85.4.1585
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Restoration of motor function to individuals who have had spinal cord injuries or stroke has been hampered by the lack of an interface to the peripheral nervous system. A suitable interface should provide selective stimulation of a large number of individual muscle groups with graded recruitment of force. We have developed a new neural interface, the Utah Slanted Electrode Array (USEA), that was designed to be implanted into peripheral nerves. Its goal is to provide such an interface that could be useful in rehabilitation as well as neuroscience applications. In this study, the stimulation capabilities of the USEA were evaluated in acute experiments in cat sciatic nerve. The recruitment properties and the selectivity of stimulation were examined by determining the target muscles excited by stimulation via each of the 100 electrodes in the array and using force transducers to record the force produced in these muscles. It is shown in the results that groups of up to 15 electrodes were inserted into individual fascicles. Stimulation slightly above threshold was selective to one muscle group for most individual electrodes. At higher currents, co-activation of agonist but not antagonist muscles was observed in some instances. Recruitment curves for the electrode array were broader with twitch thresholds starting at much lower currents than for cuff electrodes. In these experiments, it is also shown that certain combinations of electrode pairs, inserted into an individual fascicle, excite fiber populations with substantial overlap, whereas other pairs appear to address independent populations. We conclude that the USEA permits more selective stimulation at much lower current intensities with more graded recruitment of individual muscles than is achieved by conventional cuff electrodes.
引用
收藏
页码:1585 / 1594
页数:10
相关论文
共 40 条
[1]  
AGNEW WF, 1990, NEURAL PROSTHESIS FU, P107
[2]  
Akers J M, 1999, IEEE Trans Rehabil Eng, V7, P35, DOI 10.1109/86.750550
[3]   Peripheral nerve stimulation for restoration of motor function [J].
Bhadra, N ;
Peckham, PH .
JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 1997, 14 (05) :378-393
[4]  
Bhadra N, 1997, ANN BIOMED ENG, V25, P1017
[5]   ACUTE AND CHRONIC IMPLANTATION OF COILED WIRE INTRANEURAL ELECTRODES DURING CYCLICAL ELECTRICAL-STIMULATION [J].
BOWMAN, BR ;
ERICKSON, RC .
ANNALS OF BIOMEDICAL ENGINEERING, 1985, 13 (01) :75-93
[6]   A multielectrode array for intrafascicular recording and stimulation in sciatic nerve of cats [J].
Branner, A ;
Normann, RA .
BRAIN RESEARCH BULLETIN, 2000, 51 (04) :293-306
[7]   EFFECT OF ACTIVATION HISTORY ON TENSION PRODUCTION BY INDIVIDUAL MUSCLE UNITS [J].
BURKE, RE ;
RUDOMIN, P ;
ZAJAC, FE .
BRAIN RESEARCH, 1976, 109 (03) :515-529
[8]   Long-term biocompatibility of a miniature stimulator implanted in feline hind limb muscles [J].
Cameron, T ;
Liinamaa, TL ;
Loeb, GE ;
Richmond, FJR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1998, 45 (08) :1024-1035
[9]   Effects of regional stimulation using a miniature stimulator implanted in feline posterior biceps femoris [J].
Cameron, T ;
Richmond, FJR ;
Loeb, GE .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1998, 45 (08) :1036-1043
[10]   Comparison of discomfort associated with surface and percutaneous intramuscular electrical stimulation for persons with chronic hemiplegia [J].
Chae, J ;
Hart, R .
AMERICAN JOURNAL OF PHYSICAL MEDICINE & REHABILITATION, 1998, 77 (06) :516-522