Method for determination of methyl tert-butyl ether in gasoline by gas chromatography

被引:19
作者
Achten, C [1 ]
Püttmann, W [1 ]
机构
[1] Univ Frankfurt, FB Geowissensch, Inst Mineral Umweltanalyt, D-60054 Frankfurt, Germany
关键词
gasoline; methyl tert-butyl ether; ethyl tert-butyl ether; ethers; methylpentanes;
D O I
10.1016/S0021-9673(00)01220-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple method for the determination of methyl tert-butyl ether (MTBE) in gasoline has been developed. The separation of MTBE from other analytes was controlled by the use of gas chromatography-mass spectrometry in the full scan mode using the characteristic primary, secondary and tertiary ions m/z 73, 57 and 43. The sample mass spectrum did not show any superimposition of other analytes. The separation from the common gasoline component 2-methylpentane was sufficient for reliable quantitation. An application of the developed conditions using gas chromatography with flame ionization detection was performed by the analysis of regular, euro super, super premium unleaded and 'Optimax' gasoline from petrol stations in the area of Frankfurt/Main, Germany. Regular unleaded gasoline shows an average MTBE content of 0.4% (w/w), whereas the MTBE content in euro super gasoline varies between 0.4 and 4.2% (w/w). The blending of MTBE to super premium has increased from 8.2% (w/w) in 1998 to 9.8% (w/w) on average in 1999. The recently introduced gasoline 'Optimax' shows an average MTBE content of 11.9% (w/w). The presented method might also be used for the analysis of other ethers, such as ethyl tert-butyl ether, which requires the use of another internal standard. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:377 / 383
页数:7
相关论文
共 26 条
[1]   Determination of methyl tert-butyl ether in surface water by use of solid-phase microextraction [J].
Achten, C ;
Püttmann, W .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2000, 34 (07) :1359-1364
[2]  
ASTM, 1989, ANN BOOK ASTM STAND
[3]  
BREUM R, 1998, OKO TEST MAGAZIN, V4, P64
[4]  
BUND WIRTSCH, AMTL MIN BUND DEUTSC
[5]   Identification and quantitation of oxygenates in gasoline ampules using Fourier transform near-infrared and Fourier transform Raman spectroscopy [J].
Choquette, SJ ;
Chesler, SN ;
Duewer, DL ;
Wang, SW ;
OHaver, TC .
ANALYTICAL CHEMISTRY, 1996, 68 (20) :3525-3533
[6]  
*CONCAWE AUT EM MA, 1998, SURV EUR GAS QUAL SU
[7]   DETERMINATION OF OXYGENATES IN GASOLINE RANGE HYDROCARBONS BY CAPILLARY COLUMN GAS-CHROMATOGRAPHY - A USERS EXPERIENCE WITH AN OXYGEN-SPECIFIC DETECTOR (OFID) [J].
DISANZO, FP .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1990, 28 (02) :73-75
[8]  
*DTSCH WISS GES ER, 1994, ZUS OTT DTSCH RAFF
[9]  
*ENV AG UK, FUEL ADD MTBE GROUND
[10]   Analysis of gasolines by FT-IR spectroscopy [J].
Fodor, GE ;
Kohl, KB ;
Mason, RL .