Migration and failure of the Mecron screw-in acetabular prosthesis

被引:7
作者
Chell, J [1 ]
Howard, PW [1 ]
机构
[1] Derbyshire Royal Infirm, Dept Trauma & Orthopaed, Derby DE1 2QY, England
关键词
mecron acetabular prosthesis failure migration;
D O I
10.1016/S0883-5403(98)80006-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Early results with the Mecron prosthesis have been variable. We report our experience with this prosthesis at medium- to long-term follow-up. At 5 to 9 years after surgery, 43 patients (49 implants) were reviewed. Radiographic measurements of superior migration, alteration in the opening angle, and thread engagement were made. The failure rate was very high, 33% having been revised or awaiting revision. Of the remainder, 80% had migrated, 86% had changes in the opening angle, and 65% had fewer threads engaged. Many have low-grade symptoms insufficient to merit revision, but the extent of loosening makes eventual revision likely. We feel this represents an example of a new design gaining widespread usage before adequate follow-up studies.
引用
收藏
页码:638 / 641
页数:4
相关论文
共 10 条
[1]   FAILURE OF THE MECRING SCREW-RING ACETABULAR COMPONENT IN TOTAL HIP-ARTHROPLASTY - A 3-YEAR TO 7-YEAR FOLLOW-UP-STUDY [J].
BRUIJN, JD ;
SEELEN, JL ;
FEENSTRA, RM ;
HANSEN, BE ;
BERNOSKI, FP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (05) :760-766
[2]   FAILURE OF THE MECRON SCREW-IN RING [J].
CAPELLO, WN ;
COLYER, RA ;
KERNEK, CB ;
CARNAHAN, JV ;
HESS, JJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (05) :835-836
[3]   CEMENTLESS ACETABULAR COMPONENTS [J].
ENGH, CA ;
GRIFFIN, WL ;
MARX, CL .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (01) :53-59
[4]   HIP-REPLACEMENT WITH A THREADED ACETABULAR CUP - A FOLLOW-UP-STUDY [J].
FOX, GM ;
MCBEATH, AA ;
HEINER, JP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (02) :195-201
[5]  
KEATING EM, 1989, ORTHOPEDICS, V12, P1213
[6]  
KODY MH, 1990, CLIN ORTHOP RELAT R, P146
[7]  
Mittelmeier H, 1984, Hip, P312
[8]   MECHANICAL LOOSENING OF TOTAL HIP PROSTHESES - A RADIOGRAPHIC AND ROENTGEN STEREOPHOTOGRAMMETRIC STUDY [J].
MJOBERG, B ;
SELVIK, G ;
HANSSON, LI ;
ROSENQVIST, R ;
ONNERFALT, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (05) :770-774
[9]   PRIMARY MIGRATION OF FULLY-THREADED ACETABULAR PROSTHESES - A ROENTGEN STEREOPHOTOGRAMMETRIC ANALYSIS [J].
SNORRASON, F ;
KARRHOLM, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (04) :647-652
[10]  
TOOKE SM, 1988, CLIN ORTHOP RELAT R, P253