Digital automated RSA compared to manually operated RSA

被引:72
作者
Valstar, ER
Vrooman, HA
Toksvig-Larsen, S
Ryd, L
Nelissen, RGHH
机构
[1] Leiden Univ, Med Ctr, Dept Orthopaed, NL-2300 RC Leiden, Netherlands
[2] Leiden Univ, Med Ctr, Dept Radiol, Div Image Proc, Leiden, Netherlands
[3] Univ Lund Hosp, Dept Orthopaed, S-22185 Lund, Sweden
关键词
digital RSA; micromotion; accuracy; validation;
D O I
10.1016/S0021-9290(00)00138-X
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
The accuracy of digital Roentgen stereophotogrammetric analysis (RSA) was compared to the accuracy of a manually operated RSA system. For this purpose, we used radiographs of a phantom and radiographs of patients. The radiographs of the patients consisted of double examinations of 12 patients that had a tibial osteotomy and of double examinations of 12 patients that received a total hip prosthesis. First, the radiographs were measured manually with an accurate measurement table. Subsequently, the images were digitized by a film scanner at 150 DPI and 300 DPI resolutions and analyzed with the RSA-CMS software. In the phantom experiment, the manually operated system produced significantly better results than the digital system, although the maximum difference between the median values of the manually operated system and the digital system was as low as 0.013 mm for translations and 0.033 degrees for rotations. In the radiographs of the patients, the manually operated system and the digital system produced equally accurate results: no significant differences in translations and rotations were found. We conclude that digital RSA is an accurate, fast, and user friendly alternative for manually operated RSA. Currently, digital RSA systems are being used in a growing number of clinical RSA-studies. (C) 2000 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1593 / 1599
页数:7
相关论文
共 8 条
[1]
BORLIN N, 1997, THESIS UMEA U
[2]
USE OF HOUGH TRANSFORMATION TO DETECT LINES AND CURVES IN PICTURES [J].
DUDA, RO ;
HART, PE .
COMMUNICATIONS OF THE ACM, 1972, 15 (01) :11-&
[3]
Ilchmann T, 1992, J Arthroplasty, V7, P121, DOI 10.1016/0883-5403(92)90004-A
[4]
ROENTGEN STEREOPHOTOGRAMMETRY - REVIEW OF ORTHOPEDIC APPLICATIONS [J].
KARRHOLM, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1989, 60 (04) :491-503
[5]
Roentgen stereophotogrammetric analysis using computer-based image-analysis [J].
Ostgaard, SE ;
Gottlieb, L ;
ToksvigLarsen, S ;
Lebech, A ;
Talbot, A ;
Lund, B .
JOURNAL OF BIOMECHANICS, 1997, 30 (09) :993-995
[6]
RYD L, 1986, ACTA ORTHOP SCA S220, V57, P1
[7]
Selvik G., 1989, ACTA ORTHOP SCAND S, V60, P232
[8]
Fast and accurate automated measurements in digitized stereophotogrammetric radiographs [J].
Vrooman, HA ;
Valstar, ER ;
Brand, GJ ;
Admiraal, DR ;
Rozing, PM ;
Reiber, JHC .
JOURNAL OF BIOMECHANICS, 1998, 31 (05) :491-498