SQUID detected NMR and NQR

被引:22
作者
Augustine, MP [1 ]
TonThat, DM
Clarke, J
机构
[1] Univ Calif Berkeley, Dept Chem, Berkeley, CA 94720 USA
[2] Univ Calif Berkeley, Dept Phys, Berkeley, CA 94720 USA
[3] Univ Calif Berkeley, Lawrence Berkeley Lab, Dept Mat Sci, Berkeley, CA 94720 USA
基金
美国国家科学基金会;
关键词
dc SQUID; Optical Pumping; NMR; NQR; spin-echo;
D O I
10.1016/S0926-2040(97)00103-3
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The dc Superconducting QUantum Interference Device (SQUID) is a sensitive detector of magnetic flux, with a typical flux noise of the order 1 mu Phi(0) Hz(-1/2) at liquid helium temperatures. Here Phi(0) = (h) over bar/2e is the flux quantum. In our NMR or NQR spectrometer, a niobium wire coil wrapped around the sample is coupled to a thin film superconducting coil deposited on the SQUID to form a flux transformer. With this untuned input circuit the SQUID measures the flux, rather than the rate of change of flux, and thus retains its high sensitivity down to arbitrarily low frequencies. This feature is exploited in a cw spectrometer that monitors the change in the static magnetization of a sample induced by radio frequency irradiation. Examples of this technique are the detection of NQR in Al-27 in sapphire and B-11 in boron nitride, and a level crossing technique to enhance the signal of N-14 in peptides. Research is now focused on a SQUID-based spectrometer for pulsed NQR and NMR, which has a bandwidth of 0-5 MHz. This spectrometer is used with spin-echo techniques to measure the NQR longitudinal and transverse relaxation times of N-14 in NH4ClO4, 63 +/- 6 ms and 22 +/- 2 ms, respectively. With the aid of two-frequency pulses to excite the 359 kHz and 714 kHz resonances in ruby simultaneously, it is possible to obtain a two-dimensional NQR spectrum. As a third example, the pulsed spectrometer is used to study NMR spectrum of Xe-129 after polariza-tion with optically pumped Rb. The NMR line can be detected at frequencies as low as 200 Hz. At fields below about 2 mT the longitudinal relaxation time saturates at about 2000 s. Two recent experiments in other laboratories have extended these pulsed NMR techniques to higher temperatures and smaller samples. In the first, images were obtained of mineral oil floating on water at room temperature. In the second, a SQUID configured as a thin film gradiometer was used to detect NMR in a 50 mu m particle of Pt-195 at 6 mT and 4.2 K. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:139 / 156
页数:18
相关论文
共 73 条
[1]  
ABRAGAM A, 1989, PRINCIPLES NUCL MAGN, P354
[2]   Optical pumping magnetic resonance in high magnetic fields: Characterization of nuclear relaxation during pumping [J].
Augustine, MP ;
Zilm, KW .
JOURNAL OF CHEMICAL PHYSICS, 1996, 105 (08) :2998-3011
[3]   NUCLEAR MAGNETIC-RESONANCE OF N-14 AND CL-35 IN AMMONIUM-PERCHLORATE [J].
BASTOW, TJ ;
STUART, SN .
JOURNAL OF PHYSICS-CONDENSED MATTER, 1989, 1 (28) :4649-4657
[4]   Anisotropy of the proton T-1 and the low-field relaxation in NH4ClO4 below 20 K [J].
Birczynski, A ;
Lalowicz, ZT ;
Olejniczak, Z ;
Punkkinen, M .
SOLID STATE NUCLEAR MAGNETIC RESONANCE, 1996, 7 (02) :105-119
[5]   Very-Low-Temperature Static Magnetic Properties of Certain Metals Possibly Useful as Magnetic Thermometers [J].
Bishop, J. H. ;
Hirschkoff, E. C. ;
Wheatley, J. C. .
JOURNAL OF LOW TEMPERATURE PHYSICS, 1971, 5 (06) :607-616
[6]   MAGNETIC MICROSCOPY USING A LIQUID-NITROGEN COOLED YBA2CU3O7 SUPERCONDUCTING QUANTUM INTERFERENCE DEVICE [J].
BLACK, RC ;
MATHAI, A ;
WELLSTOOD, FC ;
DANTSKER, E ;
MIKLICH, AH ;
NEMETH, DT ;
KINGSTON, JJ ;
CLARKE, J .
APPLIED PHYSICS LETTERS, 1993, 62 (17) :2128-2130
[7]   HIGH-SENSITIVITY MAGNETIC-RESONANCE BY SQUID DETECTION [J].
CHAMBERLIN, RV ;
MOBERLY, LA ;
SYMKO, OG .
JOURNAL OF LOW TEMPERATURE PHYSICS, 1979, 35 (3-4) :337-347
[8]   CRYSTAL-STRUCTURE OF NH4CIO4 AT 298, 78, AND 10 DEGREESK BY NEUTRON-DIFFRACTION [J].
CHOI, CS ;
PRASK, HJ ;
PRINCE, E .
JOURNAL OF CHEMICAL PHYSICS, 1974, 61 (09) :3523-3529
[9]   BORON-NITRIDE NANOTUBES [J].
CHOPRA, NG ;
LUYKEN, RJ ;
CHERREY, K ;
CRESPI, VH ;
COHEN, ML ;
LOUIE, SG ;
ZETTL, A .
SCIENCE, 1995, 269 (5226) :966-967
[10]  
CLARKE J, 1993, SUPERCONDUCTING ELEC, P123, DOI DOI 10.1007/978-94-011-1918-4_5