Preclinical assessment of the long-term endurance of cemented hip stems.: Part 2:: in-vitro and ex-vivo fatigue damage of the cement mantle

被引:19
作者
Cristofolini, L.
Erani, P.
Savigni, P.
Bordini, B.
Viceconti, M.
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Bologna, Dept Mech Engn, Fac Engn, I-40126 Bologna, Italy
关键词
hip prostheses; cemented stem; long-term stability; fatigue testing; physiological loading; ex-vivo retrievals; fatigue cracks; micromotions; validation;
D O I
10.1243/09544119JEIM184
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Fatigue damage in the cement mantle surrounding hip stems has been studied in the past. However, so far no quantitative method has been validated for assessing ex-vivo, damage and for predicting the in-vitro risk of cement fracture. This work presents a method for measuring cement damage; the cement mantle was sliced and sections were inspected with dye penetrants and an optical microscope. Cracks were counted, measured, and classified by type in each region of the cement mantle. Statistical indicators (in total and per unit volume of cement) were proposed that allow quantitative comparison. The method was first validated on two implant types with known clinical success rate, which were tested in vitro using a physiological loading profile (described in Part I of this work). The most relevant indicators were able to detect statistical differences between the two designs. Retrieved cement mantles (the same design as one of the in-vitro stems) from revision surgery were also processed with the same inspection method. Excellent qualitative and quantitative agreement was found between the in-vitro generated fatigue damage and the cracking pattern found in the ex-vivo retrieved cement mantles. This demonstrated the effectiveness of the cement inspection protocol and provided a further validation to the in-vitro testing method.
引用
收藏
页码:585 / 599
页数:15
相关论文
共 39 条
[1]
TRANSIENT AND RESIDUAL-STRESSES AND DISPLACEMENTS IN SELF-CURING BONE-CEMENT .2. THERMOELASTIC ANALYSIS OF THE STEM FIXATION SYSTEM [J].
AHMED, AM ;
NAIR, R ;
BURKE, DL ;
MILLER, J .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1982, 104 (01) :28-37
[2]
[Anonymous], 1983, J ORTHOP RES, V1, P189
[3]
Armitage P., 2001, STAT METHODS MED RES, V4th
[4]
Baleani M, 2000, J BIOMED MATER RES, V50, P605, DOI 10.1002/(SICI)1097-4636(20000615)50:4<605::AID-JBM17>3.0.CO
[5]
2-P
[6]
Mechanical simulation of muscle loading on the proximal femur: analysis of cemented femoral component migration with and without muscle loading [J].
Britton, JR ;
Walsh, LA ;
Prendergast, PJ .
CLINICAL BIOMECHANICS, 2003, 18 (07) :637-646
[7]
Preclinical assessment of the long-term endurance of cemented hip stems. Part 1: effect of daily activities a comparison of two load histories [J].
Cristofolini, L. ;
Teutonico, A. Saponara ;
Savigni, P. ;
Erani, P. ;
Viceconti, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H6) :569-584
[8]
Comparative in vitro study on the long term performance of cemented hip stems: validation of a protocol to discriminate between "good" and "bad" designs [J].
Cristofolini, L ;
Teutonico, AS ;
Monti, L ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 2003, 36 (11) :1603-1615
[9]
CRISTOFOLINI L, 2002, P 13 C EUR SOC BIOM, P48
[10]
Culleton P, 1993, Clin Mater, V12, P95