Evaluation of the analytical performance of the Boehringer Mannheim Elecsys® 2010 immunoanalyzer

被引:60
作者
Forest, JC
Massé, J
Lane, A
机构
[1] CHU Quebec, Res Ctr, Quebec City, PQ G1L 3L5, Canada
[2] Univ Laval, Fac Med, Dept Biochem, Quebec City, PQ G1K 7P4, Canada
[3] Boehringer Mannheim Canada, Laval, PQ, Canada
关键词
chemiluminescence immunoassay; method evaluation; thyroid stimulating hormone; troponin T; carcinoembryonic antigen; prostate specific antigen;
D O I
10.1016/S0009-9120(98)00002-2
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: We evaluated the analytical performance of the Elecsys(R) 2010 immunoanalyzer (Boehringer Mannheim Canada), which is based on a new detection technology, electrochemical luminescence. Design and Methods: We used six representative assays from the initial launch menu of the instrument: thyroid stimulating hormone (TSH), free thyroxine (FT4), troponin T, human chorionic gonadotropin (hCG), carcinoembryonic antigen (CEA), and prostate specific antigen (PSA). Within-run and between-day imprecision were evaluated using pools of human specimens at low, mid, and high concentrations. Linearity was evaluated by diluting specimens with high analyte concentrations with specimens that had a low level of this analyte. Carry over and hook effect were investigated using specimens with high concentrations of hCG. Functional sensitivity was studied by running low TSH specimens over 21 daily runs, and by comparing the scatterplot of FT4 as a function of TSH. Over 100 specimens distributed across the analytical range were analyzed with two comparison methods: ES 300(R) (Boehringer Mannheim Canada) and AxSYM(R) (Abbott Laboratories). Results: Within-run and between-day imprecisions were less than 4% and 10%, respectively, for most assays. All assays were linear over the whole analytical range. Carry over was minimal (< 0.0002%). A hook effect was present for hCG levels greater than 560,000 U/L. The functional sensitivity of the TSH assay was lower than 0.02 mIU/L. Correlation coefficients were all >0.94. Small proportional errors were observed in comparison studies for the CEA and PSA assays. Conclusions: The Elecsys 2010 system was shown to have an acceptable analytical performance for the rapid analysis of a wide variety of analytes. The hook effect observed with the hCG assay would imply that the laboratory informs the clinicians of the possibility of falsely low values in trophoblastic diseases or that all specimens with values greater than 3000 Un are reassayed after dilution. Copyright (C) 1998 The Canadian Society of Clinical Chemists.
引用
收藏
页码:81 / 88
页数:8
相关论文
共 11 条
[1]  
BLACKBURN GF, 1991, CLIN CHEM, V37, P1534
[2]   SCREENING FOR DOWN-SYNDROME DURING THE FIRST AND 2ND TRIMESTERS - IMPACT OF RISK-ESTIMATION PARAMETERS [J].
FOREST, JC ;
MASSE, J ;
ROUSSEAU, F ;
MOUTQUIN, JM ;
BRIDEAU, NA ;
BELANGER, M .
CLINICAL BIOCHEMISTRY, 1995, 28 (04) :443-449
[3]  
Haeckel R, 1986, 3 ECCLS, V2
[4]  
KROLL MH, 1993, CLIN CHEM, V39, P405
[5]  
LASKY FD, 1992, P 7 INT S QUAL CONTR, P199
[6]  
Letellier M, 1996, CLIN CHEM, V42, P1695
[7]  
*NCCLS, 1992, NCCLS TENT GUID EP5
[8]   CLINICAL REVIEW-12 - THE USE AND MISUSE OF THE SENSITIVE THYROTROPIN ASSAYS [J].
NICOLOFF, JT ;
SPENCER, CA .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 1990, 71 (03) :553-558
[9]  
PASSEY RB, 1986, EP6P NCCLS
[10]  
PASSING H, 1983, J CLIN CHEM CLIN BIO, V21, P709