Frameless stereotaxy with real-time tracking of patient head movement and retrospective patient-image registration

被引:51
作者
Ryan, MJ
Erickson, RK
Levin, DN
Pelizzari, CA
Macdonald, RL
Dohrmann, GJ
机构
[1] UNIV CHICAGO,DEPT RADIOL,CHICAGO,IL 60637
[2] UNIV CHICAGO,DEPT SURG,CHICAGO,IL 60637
[3] UNIV CHICAGO,DEPT RADIAT ONCOL,CHICAGO,IL 60637
关键词
frameless stereotaxis; magnetic resonance imaging; computer-assisted neurosurgery; brain neoplasm;
D O I
10.3171/jns.1996.85.2.0287
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The accuracy of a novel frameless stereotactic system was determined during 10 surgeries performed to resect brain tumors. An array of three charge-coupled device cameras tracked the locations of infrared light-emitting diodes on a hand-held stylus and on a reference frame attached to the patient's skull with a single bone screw. Patient-image registration was achieved retrospectively by digitizing randomly chosen scalp points with the system and fitting them to a scalp surface model derived from magnetic resonance (MR) images. The reference frame enabled continual correction for patient head movements so that registration was maintained even when the patient's head was not immobilized in a surgical clamp. The location of the stylus was displayed in real-time on cross-sectional and three-dimensional MR images of the head; this information was used to predict the locations of small intracranial lesions. The average distance (and standard deviation) between the actual position of the mass and its stereotactically predicted location was 4.8 +/- 3.5 mm. The authors conclude that frameless stereotaxy can be used for accurate localization of intracranial masses without resorting to using fiducial markers during presurgical imaging and without immobilizing the patient's head during surgery.
引用
收藏
页码:287 / 292
页数:6
相关论文
共 21 条
[1]  
BARNETT GH, 1993, INTERACTIVE IMAGE GU, P113
[2]  
Bucholz RD, 1993, INTERACTIVE IMAGE GU, P179
[3]  
F Colchester at al A.C., 1994, Proceedings of Visualization in Biomedical Computing, P541
[4]  
GUTHRIE BL, 1991, PERSPECT NEUROL SURG, V2, P1
[5]  
HEILBRUN MP, 1993, INTERACTIVE IMAGE GU, P169
[6]   3-DIMENSIONAL MAGNETIC-RESONANCE IMAGES OF THE BRAIN - APPLICATION TO NEUROSURGICAL PLANNING [J].
HU, XP ;
TAN, KK ;
LEVIN, DN ;
GALHOTRA, S ;
MULLAN, JF ;
HEKMATPANAH, J ;
SPIRE, JP .
JOURNAL OF NEUROSURGERY, 1990, 72 (03) :433-440
[7]   A FRAMELESS, ARMLESS NAVIGATIONAL SYSTEM FOR COMPUTER-ASSISTED NEUROSURGERY [J].
KATO, A ;
YOSHIMINE, T ;
HAYAKAWA, T ;
TOMITA, Y ;
IKEDA, T ;
MITOMO, M ;
HARADA, K ;
MOGAMI, H .
JOURNAL OF NEUROSURGERY, 1991, 74 (05) :845-849
[8]   VOLUMETRIC STEREOTACTIC SURGICAL RESECTION OF INTRA-AXIAL BRAIN MASS LESIONS [J].
KELLY, PJ .
MAYO CLINIC PROCEEDINGS, 1988, 63 (12) :1186-1198
[9]   SURFACE OF THE BRAIN - 3-DIMENSIONAL MR IMAGES CREATED WITH VOLUME RENDERING [J].
LEVIN, DN ;
HU, XP ;
TAN, KK ;
GALHOTRA, S .
RADIOLOGY, 1989, 171 (01) :277-280
[10]   THE BRAIN - INTEGRATED 3-DIMENSIONAL DISPLAY OF MR AND PET IMAGES [J].
LEVIN, DN ;
HU, XP ;
TAN, KK ;
GALHOTRA, S ;
PELIZZARI, CA ;
CHEN, GTY ;
BECK, RN ;
CHEN, CT ;
COOPER, MD ;
MULLAN, JF ;
HEKMATPANAH, J ;
SPIRE, JP .
RADIOLOGY, 1989, 172 (03) :783-789