Pulse-shape discrimination techniques for correcting the effects of radiation damage in germanium coaxial detectors

被引:9
作者
Ho, WCG
Boggs, SE
Lin, RP
Slassi-Sennou, S
Madden, NW
Pehl, RH
Hull, EL
机构
[1] Univ Calif Berkeley, Space Sci Lab, Berkeley, CA 94720 USA
[2] Univ Calif Berkeley, Dept Phys, Berkeley, CA 94720 USA
[3] Univ Calif Berkeley, Lawrence Berkeley Lab, Berkeley, CA 94720 USA
[4] Indiana Univ, Cyclotron Facil, Bloomington, IN 47408 USA
基金
美国国家航空航天局;
关键词
germanium; detectors; gamma-ray; radiation damage;
D O I
10.1016/S0168-9002(98)00359-3
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Pusle-shape discrimination (PSD) techniques on current pulses from coaxial germanium detectors can significantly correct for charge losses due to hole trapping caused by radiation damage. Numerical simulations of PSD indicate that by measuring the two largest photon interaction locations and correcting the energy depositions separately, a significant recovery of the energy resolution and the Gaussian line shape of narrow lines in radiation damaged detectors can be obtained. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:507 / 514
页数:8
相关论文
共 12 条
[2]   GAMMASPHERE - CORRECTION TECHNIQUE FOR DETECTOR CHARGE TRAPPING [J].
GOULDING, FS ;
LANDIS, DA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1145-1149
[3]  
HO WCG, 1996, IEEE NUCL SCI S C RE
[4]  
HULL EL, 1993, INDIANA U CYCLOTRON, P143
[5]   EFFECTS PRODUCED BY FAST NEUTRON BOMBARDMENT OF GE(LI) GAMMA RAY DETECTORS [J].
KRANER, HW ;
CHASMAN, C ;
JONES, KW .
NUCLEAR INSTRUMENTS & METHODS, 1968, 62 (02) :173-+
[6]   The spectrometer SPI of the INTEGRAL mission [J].
Lichti, GG ;
Schonfelder, V ;
Diehl, R ;
Georgii, R ;
Kirchner, T ;
Vedrenne, G ;
Mandrou, P ;
vonBallmoos, P ;
Jean, P ;
Albernhe, F ;
Durouchoux, P ;
Cordier, B ;
Diallo, N ;
Sanchez, F ;
Leleux, P ;
Caraveo, PA ;
Teegarden, B ;
Matteson, J ;
Lin, R ;
Skinner, GK ;
Connell, P .
GAMMA-RAY AND COSMIC-RAY DETECTORS, TECHNIQUES, AND MISSIONS, 1996, 2806 :217-233
[7]   CHARGE CARRIER TRANSPORT PROPERTIES OF SEMICONDUCTOR-MATERIALS SUITABLE FOR NUCLEAR RADIATION DETECTORS [J].
OTTAVIANI, G ;
CANALI, C ;
ALBERIGIQUARANTA, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1975, NS22 (01) :192-204
[8]   Gamma-ray escape peak characteristics of radiation-damaged reverse-electrode germanium coaxial detectors [J].
Pehl, RH ;
Hull, EL ;
Madden, NW ;
Xing, JS ;
Friesel, DL .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 381 (01) :97-102
[9]   Simulations of pulse shape discrimination (PSD) techniques for background reduction in germanium detectors [J].
Philhour, B ;
Boggs, SE ;
Primbsch, JH ;
Slassi-Sennou, AS ;
Lin, RP ;
Feffer, PT ;
McBride, S ;
Goulding, FS ;
Madden, NW ;
Pehl, RH .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 403 (01) :136-150
[10]   LOW-NOISE TECHNIQUES IN DETECTORS [J].
RADEKA, V .
ANNUAL REVIEW OF NUCLEAR AND PARTICLE SCIENCE, 1988, 38 :217-277