Comparison of esophageal, rectal, axillary, bladder, tympanic, and pulmonary artery temperatures in children

被引:105
作者
Robinson, JL
Seal, RF
Spady, DW
Joffres, MR
机构
[1] Univ Alberta, Dept Pediat, Edmonton, AB, Canada
[2] Univ Alberta, Dept Anaesthesia, Edmonton, AB, Canada
关键词
D O I
10.1016/S0022-3476(98)70067-8
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Our objective was to determine the most reliable site for temperature measurement in children. In anesthetized children esophageal temperature readings were closest to those in the pulmonary artery (mean difference 0.1 degrees C +/- 0.5 degrees C compared with Genius tympanic thermometer (mean difference 0.6 degrees C +/- 1.0 degrees C), IVAC tympanic thermometer (mean difference 0.8 degrees C +/- 1.0 degrees C), rectal probe (mean difference 0.7 degrees C +/- 1.7 degrees C), bladder probe (mean difference 0.9 degrees C +/- 1.4 degrees C), and axillary probe (mean difference 1.3 degrees C +/- 1.3 degrees C).
引用
收藏
页码:553 / 556
页数:4
相关论文
共 11 条
[1]  
CORK RC, 1983, ANESTH ANALG, V62, P211
[2]   COMPARISON OF INTRACARDIAC AND INTRAVASCULAR TEMPERATURES WITH RECTAL TEMPERATURES IN MAN [J].
EICHNA, LW ;
BERGER, AR ;
RADER, B ;
BECKER, WH .
JOURNAL OF CLINICAL INVESTIGATION, 1951, 30 (04) :353-359
[3]   COMPARISON OF EAR-BASED, BLADDER, ORAL, AND AXILLARY METHODS FOR CORE TEMPERATURE-MEASUREMENT [J].
ERICKSON, RS ;
KIRKLIN, SK .
CRITICAL CARE MEDICINE, 1993, 21 (10) :1528-1534
[4]  
ERICKSON RS, 1994, HEART LUNG, V23, P181
[5]   THERMAL AND CARDIOVASCULAR CHANGES DURING 3 METHODS OF RESUSCITATION FROM MILD HYPOTHERMIA [J].
HAYWARD, JS ;
ECKERSON, JD ;
KEMNA, D .
RESUSCITATION, 1984, 11 (1-2) :21-33
[6]   CORE TEMPERATURE MEASURED IN THE AURICULAR CANAL - COMPARISON BETWEEN 4 DIFFERENT TYMPANIC THERMOMETERS [J].
JAKOBSSON, J ;
NILSSON, A ;
CARLSSON, L .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1992, 36 (08) :819-824
[7]   URINARY-BLADDER TEMPERATURE MONITORING - A NEW INDEX OF BODY CORE TEMPERATURE [J].
LILLY, JK ;
BOLAND, JP ;
ZEKAN, S .
CRITICAL CARE MEDICINE, 1980, 8 (12) :742-744
[8]   CORE TEMPERATURE-MEASUREMENT IN THE INTENSIVE-CARE UNIT [J].
NIERMAN, DM .
CRITICAL CARE MEDICINE, 1991, 19 (06) :818-823
[9]  
PRANSKY SM, 1991, CLIN PEDIATR, pS50
[10]   DO STANDARD MONITORING SITES REFLECT TRUE BRAIN TEMPERATURE WHEN PROFOUND HYPOTHERMIA IS RAPIDLY INDUCED AND REVERSED [J].
STONE, JG ;
YOUNG, WL ;
SMITH, CR ;
SOLOMON, RA ;
WALD, A ;
OSTAPKOVICH, N ;
SHREBNICK, DB .
ANESTHESIOLOGY, 1995, 82 (02) :344-351