Levels of metal ions after small- and large-diameter metal-on-metal hip arthroplasty

被引:188
作者
Clarke, MT
Lee, PTH
Arora, A
Villar, RN
机构
[1] SUNY Upstate Med Univ, Syracuse, NY 13202 USA
[2] Univ Cambridge, Cambridge Lea Hosp, BUPA, Cambridge Hip & Knee Unit, Cambridge CB4 9EL, England
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2003年 / 85B卷 / 06期
关键词
D O I
10.1302/0301-620X.85B6.14166
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Metal-on-metal (MOM) bearings for hip arthroplasty are increasing in popularity. Concern remains, however, regarding the potential toxicological effects of the metal ions which these bearings release. The serum levels of cobalt and chromium in 22 patients who had undergone MOM resurfacing arthroplasty were compared with a matched group of 22 patients who had undergone 28 mm MOM total hip arthroplasty (THA). At a median of 16 months (7 to 56) after resurfacing arthroplasty, we found the median serum levels of cobalt and chromium to be 38 nmol/l (14 to 44) and 53 nmol/l (23 to 165) respectively. These were significantly greater than the levels after 28 mm MOM THA which were 22 nmol/l (15 to 87, p = 0.021) and 19 nmol/l (2 to 58, p < 0.001) respectively. Since the upper limit for normal patients without implants is typically 5 nmol/l, both groups had significantly raised levels of metal ions. MOM bearings of large diameter, however, result in a greater systemic exposure of cobalt and chromium ions than bearings of small diameter. This may be of relevance for potential long-term side-effects. It is not known to what extent this difference is due to corrosion of the surfaces of the component or of the wear particles produced.
引用
收藏
页码:913 / 917
页数:5
相关论文
共 25 条
[1]
[Anonymous], 1989, TRACE ELEMENTS HUMAN
[2]
BLACK J, 1988, NATO ASI SER, P83
[3]
Elevated serum cobalt with metal-on-metal articulating surfaces [J].
Brodner, W ;
Bitzan, P ;
Meisinger, V ;
Kaider, A ;
GottsaunerWolf, F ;
Kotz, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (02) :316-321
[4]
Chan FW, 1999, CLIN ORTHOP RELAT R, P10
[5]
Increased chromosome translocations and aneuploidy in peripheral blood lymphocytes of patients having revision arthroplasty of the hip [J].
Doherty, AT ;
Howell, RT ;
Ellis, LA ;
Bisbinas, I ;
Learmonth, ID ;
Newson, R ;
Case, CP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (07) :1075-1081
[6]
Total hip arthroplasty with use of the metasul metal-on-metal articulation - Four to seven-year results [J].
Dorr, LD ;
Wang, ZI ;
Longjohn, DB ;
Dubois, B ;
Murken, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (06) :789-798
[7]
JACOBS JJ, 1996, CLIN ORTHOP S, V329, P256
[8]
Analysis of fluid film lubrication in artificial hip joint replacements with surfaces of high elastic modulus [J].
Jin, ZM ;
Dowson, D ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (03) :247-256
[9]
In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty [J].
Komistek, RD ;
Dennis, DA ;
Ochoa, JA ;
Haas, BD ;
Hammill, C .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (10) :1836-1841
[10]
Short-term results of the M2a-taper metal-on-metal articulation [J].
Lombardi, AV ;
Mallory, TH ;
Alexiades, MM ;
Cuckler, JM ;
Faris, PM ;
Jaffe, KA ;
Keating, EM ;
Nelson, CL ;
Ranawat, CS ;
Williams, J ;
Wixson, R ;
Hartman, JF ;
Capps, SG ;
Kefauver, CA .
JOURNAL OF ARTHROPLASTY, 2001, 16 (08) :122-128