Insertional torque and pull-out strengths of conical and cylindrical pedicle screws in cadaveric bone

被引:189
作者
Kwok, AWL
Finkelstein, JA
Woodside, T
Hearn, TC
Hu, RW
机构
[1] UNIV TORONTO,SUNNYBROOK HLTH SCI CTR,DIV ORTHOPAED,N YORK,ON M4N 3M5,CANADA
[2] UNIV TORONTO,SUNNYBROOK HLTH SCI CTR,ORTHOPAED BIOMECH RES LAB,N YORK,ON M4N 3M5,CANADA
关键词
conical or cylindrical thread profile; insertion torque; pedicle screws; pull-out strength;
D O I
10.1097/00007632-199611010-00004
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Design. Insertion torque and pull-out strengths of conical and cylindrical pedicle screws were compared in human cadaveric vertebral bodies. Objectives. To compare the performance of the conical design with the cylindrical design, and to determine whether insertional torque correlates with pull-out strength. Summary of Background Data. A tapered pedicle screw design may lessen the likelihood of implant failure. Its effect on thread purchase is not known. Previous studies of cylindrical designs on the relation between insertion torque and pull-out strength have been conducted in bovine and synthetic bone. Methods. Seventy-eight pedicles were assigned randomly to one of the following pedicle screw: Texas Scottish Rite Hospital (Sofamor-Danek, Memphis, TN), Steffee VSP (Acromed, Cleveland, OH), Diapason (Dimso, Paris, France), AO Schanz (Synthes, Paoli, PA), or Synthes USS (Synthes, Paoli, PA). Pedicle screws were inserted with a torque screwdriver. Each screw was extracted axially from the pedicle at a rate of 1.0 mm/sec until failure using an MTS machine (Bionix 858, Minneapolis, MN). Force data were recorded. Results. The conical design had the highest insertion torque. There were no significant differences in pull-out between any of the screw types. Correlation between insertional torque and pull-out strength was statistically significant only with the Texas Scottish Rite Hospital and Steffee VSP in L4 and AO Schanz in L5. Conclusions. A conical screw profile increases insertion torque, although insertional torque is not a reliable predictor of pull-out strength in cadaveric bone. Screw profile (with similar dimensions) has little effect on straight axial pull-out strengths in cadaveric bone.
引用
收藏
页码:2429 / 2434
页数:6
相关论文
共 14 条
[1]
RELATIONSHIP OF BONE STRENGTH AND BONE QUANITITY IN HEALTH DISEASE AND AGING [J].
BARTLEY, MH ;
ARNOLD, JS ;
HASLAM, RK ;
JEE, WSS .
JOURNALS OF GERONTOLOGY, 1966, 21 (04) :517-&
[2]
THE EFFECTS OF PEDICLE SCREW FIT - AN IN-VITRO STUDY [J].
BRANTLEY, AGU ;
MAYFIELD, JK ;
KOENEMAN, JB ;
CLARK, KR .
SPINE, 1994, 19 (15) :1752-1758
[3]
SCREW FIXATION IN THE HUMAN SACRUM - AN INVITRO STUDY OF THE BIOMECHANICS OF FIXATION [J].
CARLSON, GD ;
ABITBOL, JJ ;
ANDERSON, DR ;
KRAG, MH ;
KOSTUIK, JP ;
WOO, SLY ;
GARFIN, SR .
SPINE, 1992, 17 (06) :S196-S203
[4]
CORRELATIONS BETWEEN SCREW HOLE PREPARATION, TORQUE OF INSERTION, AND PULLOUT STRENGTH FOR SPINAL SCREWS [J].
DAFTARI, TK ;
HORTON, WC ;
HUTTON, WC .
JOURNAL OF SPINAL DISORDERS, 1994, 7 (02) :139-145
[5]
COMPLICATIONS OF LUMBAR SPINAL-FUSION WITH TRANSPEDICULAR INSTRUMENTATION [J].
DAVNE, SH ;
MYERS, DL .
SPINE, 1992, 17 (06) :S184-S189
[6]
TRANSPEDICULAR SCREW-ROD FIXATION OF THE LUMBAR SPINE - OPERATIVE TECHNIQUE AND OUTCOME IN 104 CASES [J].
DICKMAN, CA ;
FESSLER, RG ;
MACMILLAN, M ;
HAID, RW .
JOURNAL OF NEUROSURGERY, 1992, 77 (06) :860-870
[7]
SHEAR-STRENGTH OF TRABECULAR BONE FROM FEMUR, AND SOME FACTORS AFFECTING SHEAR-STRENGTH OF CEMENT-BONE INTERFACE [J].
HALAWA, M ;
LEE, AJC ;
LING, RSM ;
VANGALA, SS .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1978, 92 (01) :19-30
[8]
Hearn T. C., 1992, Veterinary and Comparative Orthopaedics and Traumatology, V5, P10
[9]
MAGERL F, 1983, EXTERNAL SKELETAL FI, P553
[10]
ANATOMIC ANALYSIS OF PEDICLE CORTICAL AND CANCELLOUS DIAMETER AS RELATED TO SCREW SIZE [J].
MISENHIMER, GR ;
PEEK, RD ;
WILTSE, LL ;
ROTHMAN, SLG ;
WIDELL, EH .
SPINE, 1989, 14 (04) :367-372