Hitachi Hemolytic Index correlates with HBOC-201 concentrations: Impact on suppression of analyte results

被引:9
作者
Moon-Massat, Paula F. [1 ]
Tierney, James P. [1 ]
Hock, Karl G. [2 ]
Scott, Mitchell G. [2 ]
机构
[1] Biopure Corp, Cambridge, MA 02141 USA
[2] Washington Univ, Sch Med, Div Lab & Genom Med, St Louis, MO 63110 USA
关键词
hemoglobin based oxygen carrier; interference; hemolysis; blood substitute; oxygen therapeutic;
D O I
10.1016/j.clinbiochem.2007.12.003
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: To determine the feasibility of laboratories to use an instrument's Hemolytic Index (HI) to determine if a test result would be accurate in the presence of hemoglobin-based oxygen carrier HBOC-201. Design and methods: HI values from the Roche Hitachi Modular P800 for samples containing HBOC-201 were determined. HI limits for 24 tests determined by addition of RBC lysate hemoglobin to serum and the instrument manufacturer's HI limits were compared to HI limits for reporting the same tests determined by adding HBOC-201 to plasma. Results: There is a linear relationship (R-2=0.99) between the HI value on the Modular P800 and HBOC-201 concentrations. Twenty-one of 24 HI limits for reporting results as determined by the manufacturer or adding RBC lysate to serum were more conservative or equal to the limit determined by HBOC-201 interference testing. HBOC-201 interference testing was more conservative for albumin, creatinine, and uric acid. Conclusion: For most analytes, the Modular P800 HI provides an accurate, conservative and automated method to determine whether laboratory results should be reported or suppressed when samples contain HBOC-201. (c) 2007 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:432 / 435
页数:4
相关论文
共 7 条
[1]  
Callas DD, 1997, CLIN CHEM, V43, P1744
[2]  
Chance JJ, 2000, CLIN CHEM, V46, P1331
[3]  
*CLSI, 2002, EP7A CLSI, V22
[4]  
Ma ZM, 1997, CLIN CHEM, V43, P1732
[5]  
*US DEP HHS, 1992, FED REGISTER, V57, P188
[6]  
Vermeer HJ, 2005, CLIN CHEM, V51, P244
[7]   Effect of the hemoglobin-based oxygen carrier HBOC-201 on laboratory instrumentation:: Cobas integra, chiron blood gas analyzer 840, Sysmex™SE-9000 and BCT [J].
Wolthuis, A ;
Peek, D ;
Scholten, R ;
Moreira, P ;
Gawryl, M ;
Clark, T ;
Westerhuis, L .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 1999, 37 (01) :71-76