Calculation of computed tomography dose index to effective dose conversion factors based on measurement of the dose profile along the fan shaped beam

被引:18
作者
Jansen, JTM
Geleijns, J
Zweers, D
Schultz, FW
Zoetelief, J
机构
[1] LEIDEN UNIV HOSP,DEPT CLIN ONCOL,2300 RC LEIDEN,NETHERLANDS
[2] LEIDEN UNIV HOSP,DEPT DIAGNOST RADIOL,2300 RC LEIDEN,NETHERLANDS
[3] TNO,ME RADIOL SERV,CTR RADIOL PROTECT & DOSIMETRY,6800 ES ARNHEM,NETHERLANDS
关键词
D O I
10.1259/0007-1285-69-817-33
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The variation in computed tomography dose index (CTDI) to effective dose conversion factors between different types of CT scanner is large (i.e. a factor of about 2 due to differences in beam shaping filters). Consequently, scanner specific conversion factors have to be applied. For some types of scanner, however, detailed information on the construction of beam shaping filters is not provided by the manufacturers. It is of interest to investigate the use of measured dose profiles for the calculation of conversion factors. Based upon measured dose profiles, two appropriate photon spectra selected on the basis of measured half value layers, gender specific adult phantoms Adam and Eva, and the Monte Carlo neutron and photon radiation transport code (MCNP), organ and effective dose conversion factors are calculated. To validate the method, a comparison is made between results for measured and calculated beam profiles for a Philips Tomoscan 350. The results in terms of effective dose per slice per unit of CTDI are compared with published data. Relative difference in conversion factors per slice averaged over all slices used for the calculations is 13+/-4% between the two spectra, 10.2+/-0.2% between measured and calculated beam profiles and 50+/-191% between the phantoms of different gender. The relative difference between the averaged results for the Adam and Eva phantoms and published results for a hermaphrodite phantom is on average equal to or less than 15+/-13%, depending on the spectrum and beam profile used, although larger differences can occur for specific slices. It is concluded that CTDI to effective dose conversion factors can be derived on the basis of measured beam profiles.
引用
收藏
页码:33 / 41
页数:9
相关论文
共 15 条
  • [1] [Anonymous], 1964, NBS HDB, V85
  • [2] BRIESMEISTER J, 1991, MCNP4 NEWSLETTER
  • [3] BRIESMEISTER J, 1988, MCNP3B NEWSLETTER
  • [4] BRIESMEISTER JF, 1986, GENERAL MONTE CARLO
  • [5] COMPARISON OF 2 METHODS FOR ASSESSING PATIENT DOSE FROM COMPUTED-TOMOGRAPHY
    GELEIJNS, J
    VANUNNIK, JG
    ZOETELIEF, J
    ZWEERS, D
    BROERSE, JJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (796) : 360 - 365
  • [6] GELEIJNS J, 1995, PATIENT DOSIMETRY DI
  • [7] Jones D, 1993, NRPBSR250 NAT RAD PR
  • [8] JONES DG, 1991, NRPBR250 NAT RAD PRO
  • [9] KRAMER R, 1986, GSF S885 I STRAHL GE
  • [10] CALCULATION OF DOSE CONVERSION FACTORS FOR POSTERIOR-ANTERIOR CHEST RADIOGRAPHY OF ADULTS WITH A RELATIVELY HIGH-ENERGY X-RAY-SPECTRUM
    SCHULTZ, FW
    GELEIJNS, J
    ZOETELIEF, J
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (800) : 775 - 785