A CdZnTe coplanar-grid detector array for environmental remediation

被引:42
作者
Luke, PN [1 ]
Amman, M [1 ]
Lee, JS [1 ]
Ludewigt, BA [1 ]
Yaver, H [1 ]
机构
[1] Univ Calif Berkeley, Lawrence Berkeley Lab, Berkeley, CA 94720 USA
关键词
CdZnTe; semiconductor detector; gamma-ray; detector array;
D O I
10.1016/S0168-9002(00)00876-7
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A portable, room-temperature gamma-ray spectrometer is being developed for use in environmental remediation activities. Good energy resolution and high-detection efficiency are achieved using an array of CdZnTe coplanar-grid detectors. The array is of modular design, which allows for the stacking of an arbitrary number of detectors, as well as for the easy exchange of detectors for repairs if needed. Each element of the array consists of a 1 cm(3) detector and low-power readout electronics. Each detector module including the electronics fits within a small footprint of 1.4 cm x 1.4 cm, thus allowing the detectors to be closely packed. A compliant detector-mounting scheme has been developed to protect the detectors from mechanical shock and vibration expected in field use. A prototype ? x ? array has been constructed and evaluated. The design of the detector, detector module and test results from the array are presented. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:319 / 324
页数:6
相关论文
共 6 条
[1]   Electron transport uniformity characterization of CdZnTe using alpha particles [J].
Amman, M ;
Lee, JS ;
Luke, PN .
HARD X-RAY AND GAMMA-RAY DETECTOR PHYSICS AND APPLICATIONS, 1998, 3446 :179-190
[2]   Optimization criteria for coplanar-grid detectors [J].
Amman, M ;
Luke, PN .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (03) :205-212
[3]   Coplanar grid patterns and their effect on energy resolution of CdZnTe detectors [J].
He, Z ;
Knoll, GF ;
Wehe, DK ;
Du, YF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 411 (01) :107-113
[4]   UNIPOLAR CHARGE SENSING WITH COPLANAR ELECTRODES - APPLICATION TO SEMICONDUCTOR-DETECTORS [J].
LUKE, PN .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :207-213
[5]   Performance of CdZnTe coplanar-grid gamma-ray detectors [J].
Luke, PN ;
Eissler, EE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1481-1486
[6]  
PRETTYMAN TH, COMMUNICATION