Determination of anionic feed additives by two-dimensional capillary isotachophoresis

被引:16
作者
Blatny, P [1 ]
Kvasnicka, F [1 ]
Kenndler, E [1 ]
机构
[1] INST CHEM TECHNOL PRAGUE,DEPT CARBOHYDRATE CHEM & TECHNOL,CZ-16000 PRAGUE,CZECH REPUBLIC
关键词
feed additives; benzoic acid; citric acid; formic acid; fumaric acid; lactic acid; malic acid; phosphoric acid; tartaric acid; organic acids;
D O I
10.1016/0021-9673(95)01380-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An isotachophoretic method was developed for the determination of organic and inorganic acids (benzoic, citric, formic, fumaric, lactic, malic, phosphoric and tartaric acid) in feed additive preparations. The resulting isotachophoretic measurements were carried out two-dimensionally in a column-coupling instrument equipped with contact conductivity (in both capillaries) and UV absorbance (in the analytical capillary) detectors, respectively. Preseparation was run in the first capillary at pH 6.1 of the leading electrolyte. The pH of the leading electrolyte of the second capillary was 2.5, with beta-cyclodextrin added to improve the separation selectivity for tartrate and fumarate. The terminating electrolyte was 5 mmol l(-1) caproic acid in all cases. The precision of the determinations including sample pretreatment, expressed as the relative standard deviation, varied between 3.3 and 7.9%, depending on the concentration levels of the analytes. The accuracy, determined by the recovery of analytes added, was typically in the order of 94 to 98%.