MR thermometry-based feedback control of laser interstitial thermal therapy at 980 nm

被引:166
作者
McNichols, RJ
Gowda, A
Kangasniemi, M
Bankson, JA
Price, RE
Hazle, JD
机构
[1] BioTex Inc, Houston, TX 77054 USA
[2] Univ Texas, MD Anderson Canc Ctr, Dept Imaging Phys, Houston, TX 77030 USA
[3] Univ Texas, MD Anderson Canc Ctr, Dept Vet Med & Surg, Houston, TX 77030 USA
关键词
brain; diode laser; LITT; magnetic resonance imaging; thermal imaging;
D O I
10.1002/lsm.10243
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Background and Objectives: The goal of this study was to explore the feasibility of magnetic resonance thermal imaging (MRTI)-based feedback control of intracerebral laser interstitial thermal therapy (LITT), using a computer workstation and 980-nm diode laser interfaced to an MR scanner. Study Design/Materials and Methods: A computer-controlled laser thermal therapy system was used to produce 12 ex vivo lesions in 3 canine and porcine brains and 16 in vivo lesions in 6 canines with diffusing tip fiberoptic applicators and energies from 54 to 900 J. MRTI predictions of thermal damage were correlated with histopathologic analysis. Results: Under feedback control, no carbonization, vaporization, or applicator damage was observed. MRTI-based prediction of thermal dose was not significantly different from histological evaluation of achieved thermal necrosis. Conclusions: The computer-controlled thermal therapy system was effective at regulating heating, eliminating carbonization and vaporization, and protecting fiberoptic applicators. MRTI estimation of thermal dose accurately predicted achieved thermal necrosis. (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:48 / 55
页数:8
相关论文
共 31 条
[1]   Novel laser system and laser irradiation method reduced the risk of carbonization during laser interstitial thermotherapy: Assessed by MR temperature measurement [J].
Atsumi, H ;
Matsumae, M ;
Kaneda, M ;
Muro, I ;
Mamata, Y ;
Komiya, T ;
Tsugu, A ;
Tsugane, R .
LASERS IN SURGERY AND MEDICINE, 2001, 29 (02) :108-117
[2]  
BROWN SG, 1983, WORLD J SURG, V7, P700
[3]   Improved quantitative MR thermometry using a 1.5 T scanner to monitor cooled applicator systems during laser-induced interstitial thermotherapy (LITT) [J].
Busse, H ;
Rassek, M ;
Grust, A ;
Kahn, T ;
Schwarzmaier, HJ .
THERMAL TREATMENT OF TISSUE WITH IMAGE GUIDANCE, PROCEEDINGS OF, 1999, 3594 :196-203
[4]   Changes in relative light fluence measured during laser heating: implications for optical monitoring and modelling of interstitial laser photocoagulation [J].
Chin, LCL ;
Whelan, WM ;
Sherar, MD ;
Vitkin, IA .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (09) :2407-2420
[5]  
DEPOORTER J, 1995, MAGN RESON MED, V34, P359, DOI 10.1002/mrm.1910340313
[6]  
DEPOORTER J, 1995, MAGNET RESON MED, V33, P74, DOI 10.1002/mrm.1910330111
[7]  
DIDERICH C, 2001, P SOC PHOTO-OPT INS, V4247, P166
[8]   Laser-induced interstitial thermotherapy of liver metastases in an interventional 0.5 tesla MRI system:: Technique and first clinical experiences [J].
Fiedler, VU ;
Schwarzmaier, HJ ;
Eickmeyer, F ;
Müller, FP ;
Schoepp, C ;
Verreet, PR .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 13 (05) :729-737
[9]  
Graham SJ, 1999, MAGNET RESON MED, V41, P321, DOI 10.1002/(SICI)1522-2594(199902)41:2<321::AID-MRM16>3.0.CO
[10]  
2-9