Biophysical characteristics of HIMAC clinical irradiation system for heavy-ion radiation therapy

被引:751
作者
Kanai, T
Endo, M
Minohara, S
Miyahara, N
Koyama-Ito, H
Tomura, H
Matsufuji, N
Futami, Y
Fukumura, A
Hiraoka, T
Furusawa, Y
Ando, K
Suzuki, M
Soga, F
Kawachi, K
机构
[1] Natl Inst Radiat Sci, Res Ctr Charged Particle Therapy, Div Accelerator Phys & Engn, Inage Ku, Chiba 263, Japan
[2] Natl Inst Radiat Sci, Med Phys & Engn Off, Chiba 263, Japan
[3] Natl Inst Radiat Sci, Div Radiat Res, Chiba 263, Japan
[4] Natl Inst Radiat Sci, Space & Particle Radiat Sci Res Grp, Chiba 263, Japan
[5] Natl Inst Radiat Sci, Div Planning & Coordinat, Chiba 263, Japan
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1999年 / 44卷 / 01期
关键词
heavy ion radiotherapy; irradiation system;
D O I
10.1016/S0360-3016(98)00544-6
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The irradiation system and biophysical characteristic carbon beams are examined regarding radiation therapy. Methods and Materials: An irradiation system was developed for heavy-ion radiotherapy. Wobbler magnets and a scatterer were used for flattening the radiation field. A patient-positioning system using X ray and image intensifiers was also installed in the irradiation system. The depth-dose distributions of the carbon beams were modified to make a spread-out Bragg peak, which was designed based on the biophysical characteristics of monoenergetic beams. A dosimetry system for heavy-ion radiotherapy was established to deliver heavy-ion doses safely to the patients according to the treatment planning. A carbon beam of 80 keV/mu m in the spread-out Bragg peak was found to be equivalent in biological responses to the neutron beam that is produced at cyclotron facility in National Institute Radiological Sciences (NIRS) by bombarding 30-MeV deuteron beam on beryllium target. The fractionation schedule of the NIRS neutron therapy was adapted for the first clinical trials using carbon beams. Results: Carbon beams, 290, 350, and 400 MeV/u, were used for a clinical trial from June of 1994. Over 300 patients have already been treated by this irradiation system by the end of 1997. (C) 1999 Elsevier Science Inc.
引用
收藏
页码:201 / 210
页数:10
相关论文
共 30 条
[1]  
ANDO K, 1993, NIRS PUBLICATION
[2]  
ANDO K, 1992, NIRS PUBLICATION
[3]  
[Anonymous], JPN J MED PHYS
[4]   INSTRUMENTATION FOR TREATMENT OF CANCER USING PROTON AND LIGHT-ION BEAMS [J].
CHU, WT ;
LUDEWIGT, BA ;
RENNER, TR .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (08) :2055-2122
[5]  
Endo M, 1996, J JASTRO, V8, P231, DOI [10.11182/jastro1989.8.231, DOI 10.11182/JASTRO1989.8.231]
[6]   Response of mouse intestine after single and fractionated irradiation with accelerated carbon ions with a spread-out Bragg peak [J].
Fukutsu, K ;
Kanai, T ;
Furusawa, Y ;
Ando, K .
RADIATION RESEARCH, 1997, 148 (02) :168-174
[7]  
FURUSAWA Y, 1993, NIRS PUBLICATION
[8]  
FUTAMI Y, 1996, JAERI C, P442
[9]  
HIRAO Y, 1992, NUCL PHYS A, V538, pC541, DOI 10.1016/0375-9474(92)90803-R
[10]  