共 3 条
Corrosion of tungsten spirals. A disturbing finding
被引:26
作者:
Weill, A
Ducros, V
Cognard, C
Piotin, M
Moret, J
机构:
[1] Fdn Ophthalmol Adolphe de Rothschild, Serv Neuroradiol Intervent, F-75019 Paris, France
[2] CHU Grenoble, Hop Tronche, Biochim Lab C, Grenoble, France
来源:
INTERVENTIONAL NEURORADIOLOGY
|
1998年
/
4卷
/
04期
关键词:
D O I:
10.1177/159101999800400412
中图分类号:
R74 [神经病学与精神病学];
学科分类号:
摘要:
Very long-term (more than 30 months after treatment) skull plain films of patients treated with Mechanical Detachable Spirals "MDS" (tungsten) for intracranial aneurysm or dural fistula by venous approach, showed a decreasing level of radiopacity of the spirals suggesting that this material was resorbing with time. To date, all the aneurysms selectively treated with "MDS'" which were followed-up by angiography move than 30 months after treatment (three cases) showed recanalization. The recanalization was proportional to degree of the "corrosion".
引用
收藏
页码:337 / 340
页数:4