Laser induced breakdown spectrometry with an echelle spectrometer and intensified charge coupled device detection

被引:69
作者
Bauer, HE
Leis, F
Niemax, K
机构
[1] Inst Spectrochem & Appl Spect, D-44139 Dortmund, Germany
[2] Univ Hohenheim, Inst Phys, D-70599 Stuttgart, Germany
关键词
laser ablation; microanalysis; LIBS; echelle-spectrometer; ICCD detection;
D O I
10.1016/S0584-8547(98)00234-1
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The combination of an echelle spectrometer with a time-gated intensified charge coupled device (ICCD) camera for simultaneous measurements of spectral lines in wide ranges of the optical spectrum is presented. The analytical and diagnostic power of the echelle-ICCD combination is demonstrated by laser induced breakdown spectrometry (LIBS) of solid samples. The apparatus allows complete elemental analysis by single shots as the spectral lines of major, minor and trace constituents are measured simultaneously. The analytical precision and detection limits in single shot measurements are improved by the evaluation of many lines for each element. Further, plasma parameters, such as the plasma temperatures, can be determined from the same spectrum and, if necessary, used far normalization of the analyte line intensities. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1815 / 1825
页数:11
相关论文
共 17 条
[1]  
BECKERROSS H, UNPUB SPECTROCHIM B
[2]   Experimental determination of laser induced breakdown thresholds of metals under nanosecond Q-switched laser operation [J].
Cabalin, LM ;
Laserna, JJ .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1998, 53 (05) :723-730
[3]   Spatial and temporal dependence of lead emission in laser-induced breakdown spectroscopy [J].
Castle, BC ;
Visser, K ;
Smith, BW ;
Winefordner, JD .
APPLIED SPECTROSCOPY, 1997, 51 (07) :1017-1024
[4]   MEASUREMENT OF TRACE-ELEMENTS IN CERAMIC AND QUARTZ BY LASER ABLATION MICROWAVE-INDUCED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
CIOCAN, A ;
HIDDEMANN, L ;
UEBBING, J ;
NIEMAX, K .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (02) :273-278
[5]   Optical emission spectroscopy studies of the influence of laser ablated mass on dry inductively coupled plasma conditions [J].
Ciocan, AC ;
Mao, XL ;
Borisov, OV ;
Russo, RE .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1998, 53 (03) :463-470
[6]   MULTISIGNAL CALIBRATION IN OPTICAL-EMISSION SPECTROSCOPY [J].
DANZER, K ;
WAGNER, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 346 (6-9) :520-524
[7]   Evaluation of laser ablation optical emission spectrometry for microanalysis in aluminium samples [J].
Geertsen, C ;
Lacour, JL ;
Mauchien, P ;
Pierrard, L .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (11) :1403-1416
[8]   Charge-transfer devices in analytical instrumentation [J].
Hanley, QS ;
Earle, CW ;
Pennebaker, FM ;
Madden, SP ;
Denton, MB .
ANALYTICAL CHEMISTRY, 1996, 68 (21) :A661-A667
[9]   SIMULTANEOUS MULTIELEMENT ANALYSIS OF SOLID SAMPLES BY LASER-ABLATION MICROWAVE-INDUCED PLASMA OPTICAL-EMISSION SPECTROMETRY [J].
HIDDEMANN, L ;
UEBBING, J ;
CIOCAN, A ;
DESSENNE, O ;
NIEMAX, K .
ANALYTICA CHIMICA ACTA, 1993, 283 (01) :152-159
[10]   Effect of sampling geometry on elemental emissions in laser-induced breakdown spectroscopy [J].
Multari, RA ;
Foster, LE ;
Cremers, DA ;
Ferris, MJ .
APPLIED SPECTROSCOPY, 1996, 50 (12) :1483-1499