Feasibility of simultaneous whole-brain imaging on an integrated PET-MRI system using an enhanced 2-point Dixon attenuation correction method

被引:25
作者
Anazodo, Udunna C. [1 ,2 ]
Thiessen, Jonathan D. [1 ,2 ]
Ssali, Tracy [1 ,2 ]
Mandel, Jonathan [3 ]
Guenther, Matthias [4 ]
Butler, John [1 ]
Pavlosky, William [3 ]
Prato, Frank S. [1 ,2 ]
Thompson, R. Terry [1 ,2 ]
Lawrence, Keith S. St. [1 ,2 ]
机构
[1] Lawson Hlth Res Inst, London, ON, Canada
[2] Western Univ, Med Biophys, London, ON, Canada
[3] St Josephs Hlth Care, Diagnost Imaging, London, ON, Canada
[4] Med Image Comp MEVIS, Fraunhofer Inst, Bremen, Germany
基金
加拿大健康研究院;
关键词
PET-MRI; attenuation correction; arterial spin labeling; ASL; F-18-fluorodexoyglucose; FDG; cerebral blood flow; glucose uptake; POSITRON-EMISSION-TOMOGRAPHY; CEREBRAL-BLOOD-FLOW; AEROBIC GLYCOLYSIS; PERFUSION MRI; FDG-PET; PET/MRI; CT; SEGMENTATION; REGISTRATION; PERFORMANCE;
D O I
10.3389/fnins.2014.00434
中图分类号
Q189 [神经科学];
学科分类号
071006 [神经生物学];
摘要
Purpose: To evaluate a potential approach for improved attenuation correction (AC) of PET in simultaneous PET and MRI brain imaging, a straightforward approach that adds bone information missing on Dixon AC was explored. Methods: Bone information derived from individual T1-weighted MRI data using segmentation tools in SPM8, were added to the standard Dixon AC map. Percent relative difference between PET reconstructed with Dixon+bone and with Dixon AC maps were compared across brain regions of 13 oncology patients. The clinical potential of the improved Dixon AC was investigated by comparing relative perfusion (rCBF) measured with arterial spin labeling to relative glucose uptake (rPET(dxbone)) measured simultaneously with F-18-flurodexoyglucose in several regions across the brain. Results: A gradual increase in PET signal from center to the edge of the brain was observed in PET reconstructed with Dixon+bone. A 5-20% reduction in regional PET signals were observed in data corrected with standard Dixon AC maps. These regional underestimations of PET were either reduced or removed when Dixon+bone AC was applied. The mean relative correlation coefficient between rCBF and rPETdxbone was r = 0.53 (p < 0.001). Marked regional variations in rCBF-to-rPET correlation were observed, with the highest associations in the caudate and cingulate and the lowest in limbic structures. All findings were well matched to observations from previous studies conducted with PET data reconstructed with computed tomography derived AC maps. Conclusion: Adding bone information derived from T1-weighted MRI to Dixon AC maps can improve underestimation of PET activity in hybrid PET-MRI neuroimaging.
引用
收藏
页数:11
相关论文
共 43 条
[1]
Combined PET/MR imaging in neurology: MR-based attenuation correction implies a strong spatial bias when ignoring bone [J].
Andersen, Flemming Littrup ;
Ladefoged, Claes Nohr ;
Beyer, Thomas ;
Keller, Sune Hogild ;
Hansen, Adam Espe ;
Hojgaard, Liselotte ;
Kjaer, Andreas ;
Law, Ian ;
Holm, Soren .
NEUROIMAGE, 2014, 84 :206-216
[2]
Unified segmentation [J].
Ashburner, J ;
Friston, KJ .
NEUROIMAGE, 2005, 26 (03) :839-851
[3]
Summary Report of the First International Workshop on PET/MR Imaging, March 19-23, 2012, Tubingen, Germany [J].
Bailey, Dale L. ;
Barthel, Henryk ;
Beyer, Thomas ;
Boellaard, Ronald ;
Gueckel, Brigitte ;
Hellwig, Dirk ;
Herzog, Hans ;
Pichler, Bernd J. ;
Quick, Harald H. ;
Sabri, Osama ;
Scheffler, Klaus ;
Schlemmer, Heinz P. ;
Schwenzer, Nina F. ;
Wehrl, Hans F. .
MOLECULAR IMAGING AND BIOLOGY, 2013, 15 (04) :361-371
[4]
Comparison of regional cerebral blood flow and glucose metabolism in the normal brain: effect of aging [J].
Bentourkia, M ;
Bol, A ;
Ivanoiu, A ;
Labar, D ;
Sibomana, M ;
Coppens, A ;
Michel, C ;
Cosnard, G ;
De Volder, AG .
JOURNAL OF THE NEUROLOGICAL SCIENCES, 2000, 181 (1-2) :19-28
[5]
MR-based attenuation correction for torso-PET/MR imaging:: pitfalls in mapping MR to CT data [J].
Beyer, Thomas ;
Weigert, Markus ;
Quick, Harald H. ;
Pietrzyk, Uwe ;
Vogt, Florian ;
Palm, Christoph ;
Antoch, Gerald ;
Mueller, Stefan P. ;
Bockisch, Andreas .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 (06) :1142-1146
[6]
Normalization in PET group comparison studies - The importance of a valid reference region [J].
Borghammer, Per ;
Jonsdottir, Kristjana Yr ;
Cumming, Paul ;
Ostergaard, Karen ;
Vang, Kim ;
Ashkanian, Mahmoud ;
Vafaee, Manoucher ;
Iversen, Peter ;
Gjedde, Albert .
NEUROIMAGE, 2008, 40 (02) :529-540
[7]
PET/MRI for Neurologic Applications [J].
Catana, Ciprian ;
Drzezga, Alexander ;
Heiss, Wolf-Dieter ;
Rosen, Bruce R. .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (12) :1916-1925
[8]
MRI-Assisted PET Motion Correction for Neurologic Studies in an Integrated MR-PET Scanner [J].
Catana, Ciprian ;
Benner, Thomas ;
van der Kouwe, Andre ;
Byars, Larry ;
Hamm, Michael ;
Chonde, Daniel B. ;
Michel, Christian J. ;
El Fakhri, Georges ;
Schmand, Matthias ;
Sorensen, Gregory .
JOURNAL OF NUCLEAR MEDICINE, 2011, 52 (01) :154-161
[9]
Toward Implementing an MRI-Based PET Attenuation-Correction Method for Neurologic Studies on the MR-PET Brain Prototype [J].
Catana, Ciprian ;
van der Kouwe, Andre ;
Benner, Thomas ;
Michel, Christian J. ;
Hamm, Michael ;
Fenchel, Matthias ;
Fischl, Bruce ;
Rosen, Bruce ;
Schmand, Matthias ;
Sorensen, A. Gregory .
JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (09) :1431-1438
[10]
Regional correlation between resting state FDG PET and pCASL perfusion MRI [J].
Cha, Yoon-Hee K. ;
Jog, Mayank A. ;
Kim, Yoon-Chung ;
Chakrapani, Shruthi ;
Kraman, Stephen M. ;
Wang, Danny J. J. .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2013, 33 (12) :1909-1914