Evaluation of a commercially available microconcentric nebulizer for inductively coupled plasma mass spectrometry

被引:95
作者
Vanhaecke, F
VanHolderbeke, M
Moens, L
Dams, R
机构
[1] Laboratory of Analytical Chemistry, Ghent University, Institute for Nuclear Sciences, B-9000 Ghent
关键词
inductively coupled plasma mass spectrometry; microconcentric nebulizer; cross-flow nebulizer;
D O I
10.1039/ja9961100543
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The performance of a commercially available microconcentric nebulizer (MCN-100, CETAC Technologies, USA) operated at flow rates ranging from less than or equal to 0.001 up to 0.65 ml min(-1) was evaluated using a Perkin-Elmer Sciex ELAN 5000 ICP-mass spectrometer, The observations made were compared with those for the standard GemTip cross-bow nebulizer, Registration of signal behaviour plots (signal intensity as a function of the nebulizer gas bow rate) at different rf powers and at different sample uptake rates allowed firstly, a systematic optimization of the operation parameters, and secondly, a comparison of the signal behaviour observed when using both types of nebulizer, The stability of the MCN-100 was evaluated at different sample uptake rates and the occurrence of memory effects was checked for a number of elements, Also, the level and the behaviour of oxide and doubly charged ions was studied, Furthermore, the susceptibility to matrix effects was investigated using synthetic matrices of different origin (acid, organic and high salt content) and it was demonstrated that, generally, matrix effects observed with both nebulizers are comparable and the MCN-100 can be used with NaCl concentrations up to 4 g l(-1) without capillary blocking, Finally, it is illustrated that the MCN-100 can be applied at sample flow rates of <5 mu l min(-1), as are encountered when coupling capillary zone electrophoresis to ICP-MS for elemental speciation studies.
引用
收藏
页码:543 / 548
页数:6
相关论文
共 49 条
[1]   SIGNAL ENHANCEMENT OF ELEMENTS DUE TO THE PRESENCE OF CARBON-CONTAINING COMPOUNDS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
JAUNAULT, L ;
MAURAS, Y ;
MERMET, JM ;
DELAPORTE, T .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1497-1498
[2]   CHARACTERIZATION AND OPTIMIZATION OF A DIRECT INJECTION NEBULIZER FOR INTRODUCTION OF ORGANIC-SOLVENTS AND VOLATILE ANALYTE SPECIES INTO AN INDUCTIVELY COUPLED PLASMA [J].
AVERY, TW ;
CHAKRABARTY, C ;
THOMPSON, JJ .
APPLIED SPECTROSCOPY, 1990, 44 (10) :1690-1698
[3]   USE OF AN ULTRASONIC NEBULIZER WITH MEMBRANE DESOLVATION FOR ANALYSIS OF VOLATILE SOLVENTS BY INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
BOTTO, RI ;
ZHU, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (09) :905-912
[4]   SAMPLE INTRODUCTION TECHNIQUES FOR ATOMIC SPECTROSCOPY [J].
BROWNER, RF ;
BOORN, AW .
ANALYTICAL CHEMISTRY, 1984, 56 (07) :A875-+
[5]   AEROSOL TRANSPORT MODEL FOR ATOMIC SPECTROMETRY [J].
BROWNER, RF ;
BOORN, AW ;
SMITH, DD .
ANALYTICAL CHEMISTRY, 1982, 54 (08) :1411-1419
[6]   SAMPLE INTRODUCTION - THE ACHILLES HEEL OF ATOMIC SPECTROSCOPY [J].
BROWNER, RF ;
BOORN, AW .
ANALYTICAL CHEMISTRY, 1984, 56 (07) :A786-&
[7]   SAMPLE INTRODUCTION IN PLASMA EMISSION AND MASS-SPECTROMETRY - PLENARY LECTURE [J].
BROWNER, RF ;
ZHU, GX .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (06) :543-547
[8]  
BROWNER RF, 1996, WINT C PLASM SPECR F, P41
[9]   ELECTROTHERMAL VAPORIZATION FOR SAMPLE INTRODUCTION IN PLASMA SOURCE SPECTROMETRY [J].
CAREY, JM ;
CARUSO, JA .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1992, 23 (05) :397-439
[10]  
CLEMONS PS, 1996, WINT C PLASM SP FORT, P189