Noise detection during bradycardia pacing with a hybrid nonthoracotomy implantable cardioverter defibrillator system: Incidence and clinical significance

被引:6
作者
Rosenthal, ME [1 ]
Paskman, C [1 ]
机构
[1] Abington Mem Hosp, Div Cardiol, Abington, PA 19001 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1998年 / 21卷 / 07期
关键词
ICD malfunction; oversensing; noise detection;
D O I
10.1111/j.1540-8159.1998.tb00208.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to prospectively evaluate the incidence of noise detection during bradycardia pacing by an FDA approved hybrid nonthoracotomy ICD system. An illustrative case report which prompted this investigation is provided. Backup bradycardia pacing by tiered therapy cardioverter defibrillators has been useful in preventing postshock bradycardia and occasionally for chronic rate support in bradycardic patients. Unexplained "noise" detected by real-time telemetry has been previously described during bradycardia pacing by a device utilizing automatic gain control for sensing. Eighteen patients were prospectively evaluated for noise detection during ventricular pacing by the ICD. Real-time telemetry was analyzed with each patient: (1) supine, (2) supine with deep inspiration/expiration, (3) supine during Valsalva, and (4) during a change in position from supine to sitting. Analysis of pacing threshold and lead impedance was made in each patient. Eleven of 18 patients had noise detected on real-time telemetry during bradycardia pacing. In 10 patients this was noted during deep inspiration/expiration, in 2 during Valsalva maneuver, and in 5 with position change. There was no evidence in any patient of lead malfunction nor any difference in pacing threshold or lead impedance bet-treen patients with noise detected versus those without it. Noise detection by an approved hybrid ICD system is common and may be due to the automatic gain control which maximizes sensitivity during bradycardia pacing. This may lead to clinically significant even ts, with both suppression of bradycardia pacing and triggering of tachycardia therapy in the absence of ventricular tachyarrhythmias in pacemaker-dependent patients.
引用
收藏
页码:1380 / 1386
页数:7
相关论文
共 14 条
[1]  
AHERN TS, 1994, J AM COLL CARDIOL, V23, P1521
[2]   CLINICAL-EXPERIENCE WITH A TIERED-THERAPY, MULTIPROGRAMMABLE ANTIARRHYTHMIA DEVICE [J].
BARDY, GH ;
TROUTMAN, C ;
POOLE, JE ;
KUDENCHUK, PJ ;
DOLACK, GL ;
JOHNSON, G ;
HOFER, B .
CIRCULATION, 1992, 85 (05) :1689-1698
[3]   IMPLANTABLE TRANSVENOUS CARDIOVERTER-DEFIBRILLATORS [J].
BARDY, GH ;
HOFER, B ;
JOHNSON, G ;
KUDENCHUK, PJ ;
POOLE, JE ;
DOLACK, GL ;
GLEVA, M ;
MITCHELL, R ;
KELSO, D .
CIRCULATION, 1993, 87 (04) :1152-1168
[4]   CLINICAL INTERACTIONS BETWEEN PACEMAKERS AND AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
CALKINS, H ;
BRINKER, J ;
VELTRI, EP ;
GUARNIERI, T ;
LEVINE, JH .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :666-673
[5]   UNIQUE SENSING ERRORS IN 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
CALLANS, DJ ;
HOOK, BG ;
KLEIMAN, RB ;
MITRA, RL ;
FLORES, BT ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (04) :1135-1140
[6]   ELECTROCARDIOGRAPHICALLY DOCUMENTED UNNECESSARY, SPONTANEOUS SHOCKS IN 241 PATIENTS WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATORS [J].
GRIMM, W ;
FLORES, BF ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11) :1667-1673
[7]  
HOOK BG, 1991, J AM COLL CARDIOL, V17, P985
[8]  
KANTER HL, 1996, AICD ADV, P7
[9]   OVERSENSING DURING VENTRICULAR PACING IN PATIENTS WITH A 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR [J].
KELLY, PA ;
MANN, DE ;
DAMLE, RS ;
REITER, MJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (07) :1531-1534
[10]   ICD DATA-STORAGE - VALUE IN ARRHYTHMIA MANAGEMENT [J].
MARCHLINSKI, FE ;
GOTTLIEB, CD ;
SARTER, B ;
FINKLE, J ;
HOOK, B ;
CALLANS, D ;
SCHWARTZMAN, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03) :527-534