Real-time measurements of endogenous CO production from vascular cells using an ultrasensitive laser sensor

被引:62
作者
Morimoto, Y
Durante, W
Lancaster, DG
Klattenhoff, J
Tittel, FK
机构
[1] Natl Def Med Coll, Dept Med Engn, Tokorozawa, Saitama 3598513, Japan
[2] Rice Univ, Rice Quantum Inst, Houston, TX 77251 USA
[3] Baylor Coll Med, Dept Med, Houston, TX 77030 USA
[4] Baylor Coll Med, Dept Pharmacol, Houston, TX 77030 USA
[5] Def Sci & Technol Org, Labs 205, Salisbury, SA 5108, Australia
来源
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY | 2001年 / 280卷 / 01期
关键词
heme oxygenase; gas detection; difference frequency generation; vascular smooth muscle cells;
D O I
10.1152/ajpheart.2001.280.1.H483
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Carbon monoxide (CO) has been implicated as a biological messenger molecule analogous to nitric oxide. A compact gas sensor based on a midinfrared laser absorption spectroscopy was developed for direct and real-time measurement of trace levels (in approximate pmol) of CO release by vascular cells. The midinfrared light is generated by difference frequency mixing of two nearinfrared lasers in a nonlinear optical crystal. A strong infrared absorption line of CO (4.61 mum) is chosen for convenient CO detection without interference from other gas species. The generation of CO from cultured vascular smooth muscle cells was detected every 20 s without any chemical modification to the CO. The sensitivity of the sensor reached 6.9 pmol CO. CO synthesis was measured from untreated control cells (0.25 nmol per 10(7) cells/h), sodium nitroprusside-treated cells (0.29 nmol per 10(7) cells/h), and hemin-treated cells (0.49 nmol per 10(7) cells/h). The sensor also detected decreases in CO production after the addition of the heme oxygenase (HO) inhibitor tin protoporphyrin-IX (from 0.49 to 0.02 nmol per 10(7) cells/h) and increases after the administration of the HO substrate hemin (from 0.27 to 0.64 nmol per 10(7) cells/h). These results demonstrate that midinfrared laser absorption spectroscopy is a useful technique for the noninvasive and real-time detection of trace levels of CO from biological tissues.
引用
收藏
页码:H483 / H488
页数:6
相关论文
共 23 条
  • [1] BRUNER F, 1993, GAS CHROMATOGRAPHIC, P72
  • [2] NEW METHOD FOR DETERMINATION OF MICROSOMAL HEME OXYGENASE (EC 1.14.99.3) BASED ON QUANTITATION OF CARBON-MONOXIDE FORMATION
    CAVALLINSTAHL, E
    JONSSON, GI
    LUNDH, B
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1978, 38 (01) : 69 - 76
  • [3] VASCULAR SMOOTH-MUSCLE CELL HEME OXYGENASES GENERATE GUANYLYL CYCLASE STIMULATORY CARBON-MONOXIDE
    CHRISTODOULIDES, N
    DURANTE, W
    KROLL, MH
    SCHAFER, AI
    [J]. CIRCULATION, 1995, 91 (09) : 2306 - 2309
  • [4] HEME OXYGENASE ACTIVITY IN THE ADULT-RAT AORTA AND LIVER AS MEASURED BY CARBON-MONOXIDE FORMATION
    COOK, MN
    NAKATSU, K
    MARKS, GS
    MCLAUGHLIN, BE
    VREMAN, HJ
    STEVENSON, DK
    BRIEN, JF
    [J]. CANADIAN JOURNAL OF PHYSIOLOGY AND PHARMACOLOGY, 1995, 73 (04) : 515 - 518
  • [5] Ontogeny of heme oxygenase activity in the hippocampus, frontal cerebral cortex, and cerebellum of the guinea pig
    Cook, MN
    Marks, GS
    Vreman, HJ
    Nakatsu, K
    Stevenson, DK
    Brien, JF
    [J]. DEVELOPMENTAL BRAIN RESEARCH, 1996, 92 (01): : 18 - 23
  • [6] Durante W, 1998, INT J MOL MED, V2, P255
  • [7] PLASMIN POTENTIATES INDUCTION OF NITRIC-OXIDE SYNTHESIS BY INTERLEUKIN-1-BETA IN VASCULAR SMOOTH-MUSCLE CELLS
    DURANTE, W
    SCHINI, VB
    CATOVSKY, S
    KROLL, MH
    VANHOUTTE, PM
    SCHAFER, AI
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1993, 264 (02): : H617 - H624
  • [8] Nitric oxide induces heme oxygenase-1 gene expression and carbon monoxide production in vascular smooth muscle cells
    Durante, W
    Kroll, MH
    Christodoulides, N
    Peyton, KJ
    Schafer, AI
    [J]. CIRCULATION RESEARCH, 1997, 80 (04) : 557 - 564
  • [9] GRANICK S, 1975, J BIOL CHEM, V250, P9215
  • [10] HEME OXYGENASE ACTIVITY IN BLOOD-VESSEL HOMOGENATES AS MEASURED BY CARBON-MONOXIDE PRODUCTION
    GRUNDEMAR, L
    JOHANSSON, MB
    EKELUND, M
    HOGESTATT, ED
    [J]. ACTA PHYSIOLOGICA SCANDINAVICA, 1995, 153 (02): : 203 - 204