Sensitive and rapid method for the simultaneous quantification of the HIV-protease inhibitors indinavir, nelfinavir, ritonavir, and saquinavir in human plasma by reversed-phase liquid chromatography

被引:27
作者
Bouley, M [1 ]
Briere, C [1 ]
Padoin, C [1 ]
Petitjean, O [1 ]
Tod, M [1 ]
机构
[1] Univ Bobigny, Hop Avicenne, Dept Pharmacotoxicol, F-93009 Bobigny, France
关键词
liquid chromatography; indinavir; nelfinavir; ritonavir; saquinavir;
D O I
10.1097/00007691-200102000-00011
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A rapid, sensitive, and specific liquid chromatography method for the simultaneous determination of four protease inhibitors (indinavir, nelfinavir, ritonavir, and saquinavir) in human plasma is described. After a liquid-liquid extraction with terbutyl methyl ether and a sequential washing of the reconstituted sample with hexane, protease inhibitors are separated on a phenyl column using a simple binary mobile phase of ammonium acetate buffer:acetonitrile (48:52) (pH = 7.5) with an ultraviolet detection at 260 nm. The standard curves are linear in the range 0.025-1 mug/mL for saquinavir. 0.1-4 mug/mL for indinavir and nelfinavir, and 0.25-10 mug/mL for ritonavir, with an average recovery ranging from 79% to 99%, and with both low interday and intraday coefficients of variation (<15%). This assay is simple, rapid (15-minute interval between runs). and useful for therapeutic monitoring of the protease inhibitors on a routine basis.
引用
收藏
页码:56 / 60
页数:5
相关论文
共 19 条
  • [1] Indinavir concentrations and antiviral effect
    Acosta, EP
    Henry, K
    Baken, L
    Page, LM
    Fletcher, CV
    [J]. PHARMACOTHERAPY, 1999, 19 (06): : 708 - 712
  • [2] Pharmacokinetics and potential interactions amongst antiretroviral agents used to treat patients with HIV infection
    Barry, M
    Mulcahy, F
    Merry, C
    Gibbons, S
    Back, D
    [J]. CLINICAL PHARMACOKINETICS, 1999, 36 (04) : 289 - 304
  • [3] DELFRAISSY JF, 1999, THERAPEUTIC CARE HIV, P51
  • [4] Rapid quantification of indinavir in human plasma by high-performance liquid chromatography with ultraviolet detection
    Foisy, ML
    Sommadossi, JP
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1999, 721 (02): : 239 - 247
  • [5] Determination of saquinavir in human plasma, saliva, and cerebrospinal fluid by ion-pair high-performance liquid chromatography with ultraviolet detection
    Hoetelmans, RMW
    vanEssenberg, M
    Meenhorst, PL
    Mulder, JW
    Beijnen, JH
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1997, 698 (1-2): : 235 - 241
  • [6] Simultaneous determination of the HIV-protease inhibitors indinavir, nelfinavir, saquinavir and ritonavir in human plasma by reversed-phase high-performance liquid chromatography
    Hugen, PWH
    Verweij-van Wissen, CPWGM
    Burger, DM
    Wuis, EW
    Koopmans, PP
    Hekster, YA
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1999, 727 (1-2): : 139 - 149
  • [7] Legrand M., 2000, Journal de Pharmacie Clinique, V19, P61
  • [8] Determination of indinavir, a HIV-1 protease inhibitor, in human plasma using ion-pair reversed-phase high-performance liquid chromatography
    Li, WZ
    Coombs, RW
    Collier, AC
    Raisys, VA
    [J]. THERAPEUTIC DRUG MONITORING, 1999, 21 (03) : 360 - 366
  • [9] Determination of ritonavir, a new HIV protease inhibitor, in biological samples using reversed-phase high-performance liquid chromatography
    Marsh, KC
    Eiden, E
    McDonald, E
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1997, 704 (1-2): : 307 - 313
  • [10] Ordered accumulation of mutations in HIV protease confers resistance to ritonavir
    Molla, A
    Korneyeva, M
    Gao, Q
    Vasavanonda, S
    Schipper, PJ
    Mo, HM
    Markowitz, M
    Chernyavskiy, T
    Niu, P
    Lyons, N
    Hsu, A
    Granneman, GR
    Ho, DD
    Boucher, CAB
    Leonard, JM
    Norbeck, DW
    Kempf, DJ
    [J]. NATURE MEDICINE, 1996, 2 (07) : 760 - 766