Safety and efficacy outcome of mentor alpha-1 inflatable penile prosthesis implantation for impotence treatment

被引:105
作者
Goldstein, I
Newman, L
Baum, N
Brooks, M
Chaikin, L
Goldberg, K
McBride, A
Krane, RJ
机构
[1] Department of Urology, Boston University School of Medicine, Boston, MA
关键词
impotence; penile prosthesis; outcome assessment;
D O I
10.1016/S0022-5347(01)65058-4
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: We investigated safety and efficacy outcome pertaining to the Mentor Alpha-1,* 3-piece inflatable penile prosthesis for impotence treatment. Materials and Methods: A 2-phase, multi-institutional, large scale retrospective study, with independently analyzed medical record (phase I) and questionnaire (phase II) data from consecutive eligible patients of 7 physician investigators was performed from March to October 1993. Results: In phase I there were no morbidities of any type in 394 of the 434 patients (90.8%) (mean age 61 years, range 24 to 88) who underwent Alpha-1 implantation (mean followup 22.2 months, range 0.67 to 44.5). The risk of prosthesis malfunction (fluid leak and auto-inflation) was 2.5%. No cylinder aneurysms were reported. A total of 93.1% of Alpha-1 devices was free from explantation (4.4%) or revision surgery (2.5%) for approximately 2 years from the original implant date. Kaplan-Meier actuarial analyses revealed that cumulative survival of the Alpha-1 prostheses at 12, 24 and 36 months was 98 +/- 1%, 93 +/- 2% and 85 +/- 7% until device malfunction, and 91 +/- 2% 83 +/- 4% and 75 +/- 7% until surgical intervention (revision or explantation). In phase II 89% of the men claimed fulfilled expectations with the Alpha-1 prosthesis as impotence treatment. Satisfaction responses 80% or greater were noted with regard to intercourse ability and confidence, and device rigidity and function. Implantation did not result in greater than 80% satisfaction in partner relationships or feelings (as judged by the patient), social or work confidence, or intercourse frequency. Factors adversely affecting satisfaction included partner feelings of dissatisfaction (as judged by the patient), specific physician investigators and need for explantation/revision surgery. Conclusions: In 1 of the largest multi-institutional implant outcome studies thus far performed, safety and efficacy data concerning the Alpha-1 contemporary inflatable device were found markedly improved over earlier inflatable prostheses and now compare favorably with historical data from noninflatable rod type devices.
引用
收藏
页码:833 / 839
页数:7
相关论文
共 27 条
[1]   FRACTURE OF SEMIRIGID PENILE PROSTHESIS - A RARE COMPLICATION [J].
AGATSTEIN, EH ;
FARRER, JH ;
RAZ, S .
JOURNAL OF UROLOGY, 1986, 135 (02) :376-377
[2]  
*AM MED SYST INC, 1993, IMP INF PAT CONS PRO
[3]   ACTUARIAL SURVIVALS OF INFLATABLE AND FLEXIBLE, HINGED PENILE PROSTHESES [J].
BARRY, JM ;
GIESY, JD ;
MCDUFFIE, R .
UROLOGY, 1982, 20 (06) :605-607
[4]   LONG-TERM RESULTS WITH THE JONAS MALLEABLE PENILE PROSTHESIS [J].
BENSON, RC ;
PATTERSON, DE ;
BARRETT, DM .
JOURNAL OF UROLOGY, 1985, 134 (05) :899-901
[5]   COMPLICATIONS OF PENILE PROSTHESES IN THE SPINAL-CORD INJURY POPULATION [J].
COLLINS, KP ;
HACKLER, RH .
JOURNAL OF UROLOGY, 1988, 140 (05) :984-985
[6]   AMS MALLEABLE PENILE PROSTHESIS [J].
DORFLINGER, T ;
BRUSKEWITZ, R .
UROLOGY, 1986, 28 (06) :480-485
[7]   EARLY EXPERIENCE WITH THE 1ST PRE-CONNECTED 3-PIECE INFLATABLE PENILE PROSTHESIS - THE MENTOR ALPHA-1 [J].
GOLDSTEIN, I ;
BERTERO, EB ;
KAUFMAN, JM ;
WITTEN, FR ;
HUBBARD, JG ;
FITCH, WP ;
GELLER, RA ;
MCKAY, DL ;
KRANE, RJ ;
BORGES, FD ;
BABAYAN, RK ;
TUTTLE, JP ;
GRUBER, MB ;
HARIK, V ;
LEVENSON, S .
JOURNAL OF UROLOGY, 1993, 150 (06) :1814-1818
[8]   SILICONE-SILVER PENILE PROSTHESIS - DESCRIPTION, OPERATIVE APPROACH AND RESULTS [J].
JONAS, U ;
JACOBI, GH .
JOURNAL OF UROLOGY, 1980, 123 (06) :865-867
[9]   LONG-TERM EVALUATION OF THE INFLATABLE PENILE PROSTHESIS [J].
JOSEPH, DB ;
BRUSKEWITZ, RC ;
BENSON, RC .
JOURNAL OF UROLOGY, 1984, 131 (04) :670-673
[10]   5-YEAR FOLLOW-UP OF THE SCOTT INFLATABLE PENILE PROSTHESIS AND COMPARISON WITH SEMIRIGID PENILE PROSTHESIS [J].
KABALIN, JN ;
KESSLER, R .
JOURNAL OF UROLOGY, 1988, 140 (06) :1428-1430