NIST-IMGC comparison of gas flows below one litre per minute

被引:12
作者
Berg, RF
Cignolo, G
机构
[1] NIST, Gaithersburg, MD 20899 USA
[2] CNR, IMGC, I-10135 Turin, Italy
关键词
D O I
10.1088/0026-1394/40/4/303
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We compared the IMGC and NIST standards for small gas flows at the IMGC. The IMGC standard is a recently developed primary flow meter that extracts a large piston out of a temperature-controlled chamber. Controlling the piston's speed holds the pressure in the chamber constant. The NIST standard is a recently developed transfer standard that was calibrated against two primary standards at NIST. The first NIST primary extracts a large piston out of an oil-filled chamber in which a metal bellows is suspended. Controlling the piston's speed holds the pressure in the bellows constant. The second NIST primary is a static gravimetry method. The results include 49 nitrogen flow rates from 0.22 mumol s(-1) to 770 mumol s(-1) (1 mumol s(-1) = 1.3448 cm(3) min(-1) of an ideal gas at the 'standard' conditions of 0 degreesC and 101 325 Pa). For all but one of the comparisons, the agreement between IMGC and NIST is better than 0.06%.