Determination of trace level bromate in drinking water by direct injection ion chromatography

被引:45
作者
Jackson, LK [1 ]
Joyce, RJ [1 ]
Laikhtman, M [1 ]
Jackson, PE [1 ]
机构
[1] Dionex Corp, Sunnyvale, CA 94088 USA
关键词
trace analysis; bromate;
D O I
10.1016/S0021-9673(98)00774-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Disinfection byproduct anions such as bromate, chlorite and chlorate pose significant health risks, even at low mu g/l levels in drinking water. A direct injection, ion chromatographic method was developed using a Dionex AS9-HC anion-exchange column with a carbonate eluent and suppressed conductivity detection for the determination of these disinfection byproduct anions, and bromide, at low mu g/l levels in drinking water. No additional sample pretreatment, other than filtration, is required. The method is linear for the oxyhalides and bromide over the typical concentration range expected for these analytes in drinking water; and quantitative recoveries were obtained for drinking water samples spiked at 10 mu g/l. This ion chromatographic method, based on the recently developed AS9-HC column, is applicable for the quantitation of bromate in finished drinking water at the 10 mu g/l maximum contaminant level currently being proposed by the US EPA. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:187 / 192
页数:6
相关论文
共 19 条
[1]   Electrospray ion chromatography tandem mass spectrometry of oxyhalides at sub-ppb levels [J].
Charles, L ;
Pépin, D .
ANALYTICAL CHEMISTRY, 1998, 70 (02) :353-359
[2]   Electrospray ion chromatography tandem mass spectrometry of bromate at sub-ppb levels in water [J].
Charles, L ;
Pepin, D ;
Casetta, B .
ANALYTICAL CHEMISTRY, 1996, 68 (15) :2554-2558
[3]   Determination of bromate in the presence of brominated haloacetic acids by ion chromatography with inductively coupled plasma mass spectrometric detection [J].
Creed, JT ;
Magnuson, ML ;
Brockhoff, CA .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1997, 31 (07) :2059-2063
[4]  
*DION CORP, 1997, IONPAC AS9 HC PROD T
[5]  
Haddad PR, 1990, J CHROMATOGR LIB, V46
[6]   ANALYSIS OF OXYHALIDE DISINFECTION BY-PRODUCTS AND OTHER ANIONS OF INTEREST IN DRINKING-WATER BY ION CHROMATOGRAPHY [J].
HAUTMAN, DP ;
BOLYARD, M .
JOURNAL OF CHROMATOGRAPHY, 1992, 602 (1-2) :65-74
[7]  
HAUTMAN DP, 1997, PITTSB C EXP ATL GA
[8]  
KRUITHOF JC, 1995, WATER SUPP, V13, P117
[9]   Factors controlling ion-exchange selectivity in suppressed ion chromatography [J].
Pohl, CA ;
Stillian, JR ;
Jackson, PE .
JOURNAL OF CHROMATOGRAPHY A, 1997, 789 (1-2) :29-41
[10]  
SAINI C, 1997, INT ION CHROM S SANT