Ultraviolet absorbance and inductively coupled plasma mass spectrometric detection for capillary electrophoresis - A comparison of detection modes and interface designs

被引:57
作者
Sutton, KL [1 ]
B'Hymer, C [1 ]
Caruso, JA [1 ]
机构
[1] Univ Cincinnati, Dept Chem, Cincinnati, OH 45221 USA
关键词
capillary electrophoresis; ultraviolet absorbance detection; inductively coupled plasma mass spectrometric; detection;
D O I
10.1039/a801869c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A comparison of ICP-MS and UV absorbance as detectors for capillary electrophoresis (CE) was performed. A high efficiency nebulizer (HEN) with cyclone spray chamber was found to yield detection limits in the low parts-per-billion range while UV detection gave detection limits in the parts-per-million range for the same separation of the lanthanides. Electrokinetic and hydrostatic methods of injection were employed, and it,vas found that both methods gave similar detection limits and precision values for the separation, for both methods of detection. The use of a traditional concentric nebulizer as part of the CE-ICP-MS interface was evaluated and compared with the HEN. The optimum make-up buffer liquid how rate was evaluated for the concentric nebulizer and detection limits for the lanthanide separation were calculated. Detection limits with the concentric nebulizer were not improved compared with UV detection, illustrating that the choice of nebulizer for the interface is of primary importance for improved analytical sensitivity. Attempts to use an oscillating capillary nebulizer as part of a CE-ICP-MS interface are also described.