Effect of different location of atrial lead position on nearfield and farfield electrograms in dual chamber pacemaker-defibrillators

被引:10
作者
Kantharia, BK [1 ]
Wilbur, SL [1 ]
Padder, FA [1 ]
Pennington, JC [1 ]
Samuels, FL [1 ]
Movsowitz, CM [1 ]
Hessen, SE [1 ]
Saari, C [1 ]
Kutalek, SP [1 ]
机构
[1] Med Coll Penn & Hahnemann Univ, Div Cardiac Electrophysiol, Philadelphia, PA 19102 USA
关键词
dual chamber pacemaker-cardioverter defibrillator; farfield ventricular electrogram; farfield R wave; nearfield atrial electrogram; ventricular fibrillation;
D O I
10.1023/A:1009805807837
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The normal functioning of dual chamber pacemaker-cardioverter defibrillator (AV pacer/ICD) may be affected by oversensing of the farfield R wave (FFRW) by the atrial channel. This study aimed to investigate whether placement of the AV pacer/ICD's atrial lead at a lateral (LAT) wall location compared to a medial (MED) location i.e. the appendage of the right atrium, would reduce the amplitude of FFRWs but not the nearfield atrial electrograms (AEGMs) during sinus rhythm (SR) and ventricular fibrillation (VF). In 17 patients, real time electrograms were recorded during SR and induced VF through the atrial lead initially at the MED and subsequently at the LAT location. In 10 patients the electrograms in SR were also recorded on a computerized data acquisition and recording system at different band-pass filter settings. Although FFRWs were recorded both at MED and LAT locations, they were much smaller, 3.5 +/-4.1[emsp4 ]mm during SR and 1.7 +/-2.2 mm during VF at the LAT location. At 30-500 Hz band-pass filter, lower amplitudes of FFRWs 0.14 +/-0.09 mV were recorded at the LAT location. The V/A ratios of the amplitudes of FFRWs and AEGMs were smaller at the LAT location during SR and VF. The nearfield AEGMs were of similar amplitudes at the MED and LAT locations. These data indicate that lower amplitudes of FFRWs are recorded by placement of the atrial lead at the lateral wall of the right atrium. Oversensing of FFRWs may be prevented to improve functioning of the AV pacer-ICD.
引用
收藏
页码:59 / 66
页数:8
相关论文
共 14 条
[1]   FAR-FIELD QRS COMPLEX SENSING VIA THE ATRIAL PACEMAKER LEAD .2. PREVALENCE, CLINICAL-SIGNIFICANCE AND POSSIBILITY OF INTRAOPERATIVE PREDICTION IN DDD PACING [J].
BRANDT, J ;
FAHRAEUS, T ;
SCHULLER, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1540-1544
[2]   FAR-FIELD QRS COMPLEX SENSING VIA THE ATRIAL PACEMAKER LEAD .1. MECHANISM, CONSEQUENCES, DIFFERENTIAL-DIAGNOSIS AND COUNTERMEASURES IN AAI AND VDD/DDD PACING [J].
BRANDT, J ;
FAHRAEUS, T ;
SCHULLER, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (10) :1432-1438
[3]  
De Bakker J., 1995, CARDIAC ELECTROPHYSI, P1068
[4]   MEASUREMENT OF VENTRICULAR ELECTROGRAM AMPLITUDE DURING INTRAOPERATIVE INDUCTION OF VENTRICULAR TACHYARRHYTHMIAS [J].
ELLENBOGEN, KA ;
WOOD, MA ;
STAMBLER, BS ;
WELCH, WJ ;
DAMIANO, RJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (11) :1017-1022
[5]   SENSING CHARACTERISTICS OF THE RIGHT ATRIAL APPENDAGE ELECTRODE [J].
GRIFFIN, JC .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (01) :22-25
[6]  
KAY GN, 1996, CARDIAC PACING, P37
[7]  
KAY GN, 1995, CLIN CARDIAC PACING, P38
[8]   CORRELATION BETWEEN THE VENTRICULAR ELECTROGRAM AMPLITUDE IN SINUS RHYTHM AND IN VENTRICULAR-FIBRILLATION [J].
LEITCH, JW ;
YEE, R ;
KLEIN, GJ ;
JONES, DL ;
MURDOCK, CJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (09) :1105-1109
[9]   PERMANENT PERVENOUS ATRIAL-PACING FROM CORONARY VEIN - LONG-TERM FOLLOW-UP [J].
MOSS, AJ ;
RIVERS, RJ ;
KRAMER, DH .
CIRCULATION, 1974, 49 (02) :222-225
[10]  
NOWAK B, 1998, PACE, V21, P936