Evaluation of the implantable force transducer for chronic tendon-force recordings

被引:27
作者
Herzog, W
Archambault, JM
Leonard, TR
Nguyen, HK
机构
[1] Faculty of Physical Education, University of Calgary, Calgary
基金
加拿大自然科学与工程研究理事会; 英国医学研究理事会;
关键词
D O I
10.1016/0021-9290(95)00019-4
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In vivo force measurements from tendons and ligaments have become an important technique to determine internal forces, joint loading, and mechanisms of neuromotor control. The most frequently used transducers for such force recordings were placed external to the target tissue. Because of space restrictions and the associated impingement artifacts, these external transducers cannot be used for force measurements in all tendons and ligaments. In these situations, transducers placed inside the target tissue have been used recently; however, the suitability and performance characteristics of these internal transducers have not been assessed systematically. The purpose of this study was to assess the suitability and performance characteristics of an internally placed force transducer which has been used previously. The results indicated that small angular displacements of the transducer within the target tissue, as well as small relative rotations of the corresponding bones, resulted in substantially changed transducer output for given externally applied loads. Also, the transducer output was found to depend on the rate of load application. It was concluded that, although the internal force transducer gave reliable signals within a given experiment, and thus, could be used to assess relative changes in tissue forces pre- and post-interventions, it would be difficult to use the transducer for the accurate determination of the actual tissue forces during unrestrained animal locomotion.
引用
收藏
页码:103 / 109
页数:7
相关论文
共 21 条
[1]   THE DISTAL HINDLIMB MUSCULATURE OF THE CAT - PATTERNS OF NORMAL USE [J].
ABRAHAM, LD ;
LOEB, GE .
EXPERIMENTAL BRAIN RESEARCH, 1985, 58 (03) :580-593
[2]   INVITRO LIGAMENT TENSION PATTERN IN THE FLEXED KNEE IN PASSIVE LOADING [J].
AHMED, AM ;
HYDER, A ;
BURKE, DL ;
CHAN, KH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1987, 5 (02) :217-230
[3]   DESIGN AND PERFORMANCE OF A MODIFIED BUCKLE TRANSDUCER FOR THE MEASUREMENT OF LIGAMENT TENSION [J].
BARRY, D ;
AHMED, AM .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1986, 108 (02) :149-152
[4]  
CUMMINGS JF, 1991, 37TH T ANN M ORTH RE, V16, P601
[5]   IN-VITRO EVALUATION OF AN IMPLANTABLE FORCE TRANSDUCER (IFT) IN A PATELLAR TENDON MODEL [J].
GLOS, DL ;
BUTLER, DL ;
GROOD, ES ;
LEVY, MS .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (04) :335-343
[6]  
GLOS DL, 1990, 38TH ANN M ORTH RES
[7]   MECHANICAL OUTPUT OF THE CAT SOLEUS DURING TREADMILL LOCOMOTION - INVIVO VS INSITU CHARACTERISTICS [J].
GREGOR, RJ ;
ROY, RR ;
WHITING, WC ;
LOVELY, RG ;
HODGSON, JA ;
EDGERTON, VR .
JOURNAL OF BIOMECHANICS, 1988, 21 (09) :721-732
[8]   FORCE LENGTH PROPERTIES AND FUNCTIONAL DEMANDS OF CAT GASTROCNEMIUS, SOLEUS AND PLANTARIS MUSCLES [J].
HERZOG, W ;
LEONARD, TR ;
RENAUD, JM ;
WALLACE, J ;
CHAKI, G ;
BORNEMISZA, S .
JOURNAL OF BIOMECHANICS, 1992, 25 (11) :1329-1335
[9]   FORCES IN GASTROCNEMIUS, SOLEUS, AND PLANTARIS TENDONS OF THE FREELY MOVING CAT [J].
HERZOG, W ;
LEONARD, TR ;
GUIMARAES, ACS .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :945-953
[10]   VALIDATION OF OPTIMIZATION MODELS THAT ESTIMATE THE FORCES EXERTED BY SYNERGISTIC MUSCLES [J].
HERZOG, W ;
LEONARD, TR .
JOURNAL OF BIOMECHANICS, 1991, 24 :31-39