Amperometric microbial biosensor for p-nitrophenol using Moraxella sp.-modified carbon paste electrode

被引:133
作者
Mulchandani, P [1 ]
Hangarter, CM [1 ]
Lei, Y [1 ]
Chen, W [1 ]
Mulchandani, A [1 ]
机构
[1] Univ Calif Riverside, Dept Environm Chem & Engn, Riverside, CA 92521 USA
关键词
amperometric; microbial biosensor; p-nitrophenol; Moraxella sp;
D O I
10.1016/j.bios.2004.11.011
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
An amperometric microbial biosensor for highly specific, sensitive and rapid quantitative determination of p-nitrophenol was developed. The biosensor takes advantage of the ability of Moraxella sp. to specifically degrade p-nitrophenol to hydroquinone, a more electroactive compound than p-nitrophenol. The electrochemical oxidation current of hydroquinone formed in biodegradation of p-nitrophenol was measured at Moraxella sp-modified carbon paste electrode and correlated to p-phenol concentrations. The optimum response was realized by electrode constructed using 15 mg of dry cell weight per I g of carbon paste and operating at 0.3 V (versus Ag/AgCl reference) in pH 7.5, 20 mM sodium phosphate buffer. Operating at these optimum conditions the biosensor had excellent selectivity against phenol derivatives and was able to measure as low as 20 nM (2.78 ppb) p-nitrophenol with very good accuracy and reproducibility. The biosensor was stable for approximately 3 weeks when stored at 4 degrees C. The applicability of the biosensor to measure p-nitrophenol in lake water was demonstrated. (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:523 / 527
页数:5
相关论文
共 23 条
  • [1] [Anonymous], ANAL CHEM
  • [2] DETERMINATION OF 2-NITROPHENOL, 4-NITROPHENOL, 2-METHOXY-5-NITROPHENOL, AND 2,4-DINITROPHENOL BY DIFFERENTIAL-PULSE VOLTAMMETRY AND ADSORPTIVE STRIPPING VOLTAMMETRY
    BAREK, J
    EBERTOVA, H
    MEJSTRIK, V
    ZIMA, J
    [J]. COLLECTION OF CZECHOSLOVAK CHEMICAL COMMUNICATIONS, 1994, 59 (08) : 1761 - 1771
  • [3] Cordero-Rando MD, 1999, MIKROCHIM ACTA, V132, P7
  • [4] MICROBIAL TRANSFORMATION OF NITROAROMATIC COMPOUNDS IN SEWAGE EFFLUENT
    HALLAS, LE
    ALEXANDER, M
    [J]. APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 1983, 45 (04) : 1234 - 1241
  • [5] DEGRADATION AND INDUCTION SPECIFICITY IN ACTINOMYCETES THAT DEGRADE P-NITROPHENOL
    HANNE, LF
    KIRK, LL
    APPEL, SM
    NARAYAN, AD
    BAINS, KK
    [J]. APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 1993, 59 (10) : 3505 - 3508
  • [6] VOLTAMMETRIC DETERMINATION OF METHYL PARATHION, ORTHO, META AND PARA NITROPHENOL WITH A CARBON PASTE ELECTRODE MODIFIED WITH C-18
    HERNANDEZ, L
    HERNANDEZ, P
    VICENTE, J
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 345 (11): : 712 - 715
  • [7] Voltammetric determination of 4-nitrophenol at a sodium montmorillonite-anthraquinone chemically modified glassy carbon electrode
    Hu, SS
    Xu, CL
    Wang, GP
    Cui, DF
    [J]. TALANTA, 2001, 54 (01) : 115 - 123
  • [8] BIODEGRADATION OF P-NITROPHENOL VIA 1,2,4-BENZENETRIOL BY AN ARTHROBACTER SP
    JAIN, RK
    DREISBACH, JH
    SPAIN, JC
    [J]. APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 1994, 60 (08) : 3030 - 3032
  • [9] A microbial biosensor for p-nitrophenol using Arthrobacter sp.
    Lei, Y
    Mulchandani, P
    Chen, W
    Wang, J
    Mulchandani, A
    [J]. ELECTROANALYSIS, 2003, 15 (14) : 1160 - 1164
  • [10] Biosensor for direct determination of organophosphate nerve agents using recombinant Escherichia coli with surface-expressed organophosphorus hydrolase.: 1.: Potentiometric microbial electrode
    Mulchandani, A
    Mulchandani, P
    Kaneva, I
    Chen, W
    [J]. ANALYTICAL CHEMISTRY, 1998, 70 (19) : 4140 - 4145