The TEST 1 automated system - A new method for measuring the erythrocyte sedimentation rate

被引:60
作者
Plebani, M
De Toni, S
Sanzari, MC
Bernardi, D
Stockreiter, E
机构
[1] Univ Hosp Padua, Dept Lab Med, Padua, Italy
[2] Biochem Res Ctr, Castelfranco Veneto, Italy
关键词
erythrocyte sedimentation rate; Westergren ESR method; ICSH reference method; automated analyzer;
D O I
10.1093/ajcp/110.3.334
中图分类号
R36 [病理学];
学科分类号
100104 [病理学与病理生理学];
摘要
We evaluated performance of the TEST 1 (SIRE Analytical Systems, Udine, Italy), a fully automated analyzer for the measurement of the erythrocyte sedimentation rate (ESR). Intra-assay reproducibility was satisfactory for a wide range of ESR values, whereas there was a significant decrease in ESR data when the samples were stored at 4 degrees C for up to 24 hours. We compared TEST 1 with the Westergren ESR method approved by the International Council for Standardization in Haematology (ICSH) and with the Diesse Ves-Matic analyzer: Linear regression analysis comparing the TEST 1 and the ICSH reference method yielded satisfactory correlations for K-3 EDTA- and sodium citrate-anticoagulated samples. Bland-Altman analysis showed no evidence of a systematic bias between the TEST 1 and the reference method. A close correlation was found between the TEST 1 system and the Diesse Ves-Matic analyzer despite a significant positive systematic bias. Reference values for men and women were analyzed according to nonparametric statistics. The TEST 1 was easy to use, had a satisfactory operative praticability, required minimal maintenance and reduced contact with potential biohazards. This system enables the determination of ESR with any common standard-sized tube; the use of samples anticoagulated with K-3 EDTA can widely reduce the workload in clinical laboratories.
引用
收藏
页码:334 / 340
页数:7
相关论文
共 19 条
[1]
[Anonymous], 1973, Br J Haematol, V24, P671
[2]
[Anonymous], 1921, Acta. Med. Scand, DOI [10.1152/physrev.1929.9.2.241, DOI 10.1111/J.0954-6820.1921.TB15179.X]
[3]
BAIN BJ, 1983, CLIN LAB HAEMATOL, V5, P45
[4]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]
BULL BS, 1993, J CLIN PATHOL, V46, P198
[6]
ASSESSMENT OF DIESSE VES-MATIC AUTOMATED-SYSTEM FOR MEASURING ERYTHROCYTE SEDIMENTATION-RATE [J].
CASWELL, M ;
STUART, J .
JOURNAL OF CLINICAL PATHOLOGY, 1991, 44 (11) :946-949
[7]
*DDV SOFTW, 1995, STAT ADD IN MICR EXC
[8]
GAMBINO S R, 1965, Tech Bull Regist Med Technol, V35, P1
[9]
GARDNER HJ, 1989, BRIT MED J, P77
[10]
HERTZMAN A, 1993, J RHEUMATOL, V20, P2178