Effects of high pressure in liquid chromatography

被引:171
作者
Martin, M
Guiochon, G
机构
[1] Univ Tennessee, Dept Chem, Knoxville, TN 37996 USA
[2] Ecole Super Phys & Chim Ind Ville Paris, Phys & Mecan Milieux Heterogenes Lab, F-75231 Paris, France
[3] Oak Ridge Natl Lab, Div Chem Sci, Oak Ridge, TN 37831 USA
基金
美国国家科学基金会;
关键词
pressure;
D O I
10.1016/j.chroma.2005.06.005
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
All the experimental parameters that the chromatographers are used to consider as constant (the column length and its diameter, the particle size, the column porosities, the phase ratio, the column hold-up volume, the pressure gradient along the column, the mobile phase density and its viscosity, the diffusion coefficients, the equilibrium constants, the retention factors, the efficiency parameters) depend on pressure to some extent. While this dependence is negligible as long as experiments, measurements, and separations are carried out under conventional pressures not exceeding a few tens of megapascal, it is no longer so when the inlet pressure becomes much larger and exceeds 100 MPa. Equations are developed to determine the extent of the influence of pressure on all these parameters and to account for it. The results obtained are illustrated with graphics. The essential results are that (1) many parameters depend on the inlet pressure, hence on the flow rate; (2) the apparent reproducibility of parameters as simple as the retention factor will be poor if measurements are carried out at different flow rates, unless due corrections are applied to the results; (3) the influence of the temperature on the equilibrium constants should be studied under constant inlet pressure rather than at a constant flow rate, to minimize the coupling effect of pressure and temperature through the temperature dependence of the viscosity; and (4) while reproducibility of results obtained at constant pressure and flow rate will not be affected, method development becomes far more complex because of the pressure dependence of everything. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:16 / 38
页数:23
相关论文
共 51 条
[1]  
BIDLINGMEYER BA, 1969, SEP SCI, V4, P439
[2]  
BRIDGMAN PW, 1970, PHYS HIGH PRESSURES
[3]   Pressure-induced retention of the lysozyme on reversed-phase liquid chromatography [J].
Chen, SH ;
Ho, CT ;
Hsiao, KY ;
Chen, JM .
JOURNAL OF CHROMATOGRAPHY A, 2000, 891 (02) :207-215
[4]  
COLIN H, 1980, COMMUNICATION
[5]  
Darcy H.P.G., 1856, PUBLIC FOUNTAINS CIT
[6]   THE TAIT EQUATION - 100 YEARS ON [J].
DYMOND, JH ;
MALHOTRA, R .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 1988, 9 (06) :941-951
[8]   MEASUREMENT + INTERPRETATION OF C TERMS OF GAS CHROMATOGRAPHY [J].
GIDDINGS, JC ;
SCHETTLER, PD .
ANALYTICAL CHEMISTRY, 1964, 36 (08) :1483-&
[9]   STUDY OF PERFORMANCES OF THIN-LAYER CHROMATOGRAPHY .2. BAND BROADENING AND PLATE HEIGHT EQUATION [J].
GUIOCHON, G .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1978, 16 (10) :470-481
[10]   INFLUENCE OF PRESSURE ON SOLUTE RETENTION IN LIQUID-CHROMATOGRAPHY [J].
GUIOCHON, G ;
SEPANIAK, MJ .
JOURNAL OF CHROMATOGRAPHY, 1992, 606 (02) :248-250