Holmium:Yag lithotripsy efficiency varies with energy density

被引:101
作者
Vassar, GJ [1 ]
Teichman, JMH
Glickman, RD
机构
[1] Univ Texas, Hlth Sci Ctr, Div Urol, San Antonio, TX 78284 USA
[2] Univ Texas, Hlth Sci Ctr, Dept Ophthalmol, San Antonio, TX 78284 USA
关键词
holmium; lithotripsy; fiberoptics; urinary calculi;
D O I
10.1016/S0022-5347(01)62927-6
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: We test the hypothesis that holmium:YAG lithotripsy efficiency varies with optical fiber size and energy settings (energy density). Materials and Methods: The 272, 365, 550 and 940 mu m. optical fibers delivered 1 kJ. total holmium:YAG energy to calcium oxalate monohydrate calculi at energy output/pulse of 0.2 to 1.5 J. Stone mass loss was measured for each fiber energy setting. Stone crater width was characterized for single energy pulses. Fiber energy outputs were compared before and after lithotripsy. Results: Stone mass loss correlated inversely with optical fiber diameter (p <0.05). Stone loss correlated with energy/pulse for the 365, 550 and 940 mu m. fibers (p <0.001). The 272 and 365 mu m. fibers yielded equivalent stone loss at 0.2 and 0.5 J. per pulse. At energies of 1.0 J, per pulse or greater the 272 mu m. optical fiber was prone to damage, and yielded reduced energy output and stone loss compared to the 365 mu m. fiber (p <0.01). Stone crater width for single pulse energies correlated with energy settings for all fibers (p <0.001). Conclusions: Lithotripsy efficiency with the holmium:YAG laser depends on pulse energy output and diameter of the optical delivery fiber, implying that lithotripsy efficiency correlates with energy density. The 365 mu m fiber is indicated for most lithotripsy applications. The 272 mu m. fiber is susceptible to damage and inefficient energy transmission at energies of 1.0 J. per pulse or greater. The 272 mu m. fiber is indicated at energies of less than 1.0 J. per pulse for small caliber ureteroscopes or when maximal flexible ureteroscope deflection is required.
引用
收藏
页码:471 / 476
页数:6
相关论文
共 22 条
[1]  
Aslan Peter, 1997, Journal of Endourology, V11, pS47
[2]  
BEGHUIN D, 1998, SPIE, V3195, P220
[3]   A COMPARISON OF THE EFFECTS OF ND-YAG AND HO-YAG LASER IRRADIATION ON DENTIN AND ENAMEL [J].
CERNAVIN, I .
AUSTRALIAN DENTAL JOURNAL, 1995, 40 (02) :79-84
[4]   Holmium:YAG laser-induced damage to guidewires:: Experimental study [J].
Freiha, GS ;
Glickman, RD ;
Teichman, JMH .
JOURNAL OF ENDOUROLOGY, 1997, 11 (05) :331-336
[5]  
Grant SA, 1997, LASER SURG MED, V21, P65, DOI 10.1002/(SICI)1096-9101(1997)21:1<65::AID-LSM10>3.0.CO
[6]  
2-8
[7]  
JACQUES SL, 1992, SURG CLIN N AM, V72, P531
[8]   TEMPERATURE-DEPENDENCE OF THE ABSORPTION-COEFFICIENT OF WATER FOR MIDINFRARED LASER-RADIATION [J].
JANSEN, ED ;
VANLEEUWEN, TG ;
MOTAMEDI, M ;
BORST, C ;
WELCH, AJ .
LASERS IN SURGERY AND MEDICINE, 1994, 14 (03) :258-268
[9]  
Jansen ED, 1996, LASER SURG MED, V18, P278, DOI 10.1002/(SICI)1096-9101(1996)18:3<278::AID-LSM10>3.0.CO
[10]  
2-2