Removal of phase artifacts from fMRI data using a Stockwell transform filter improves brain activity detection

被引:46
作者
Goodyear, BG
Zhu, HM
Brown, RA
Mitchell, JR
机构
[1] Foothills Med Ctr, Seaman Family MR Res Ctr, Calgary, AB T2N 2T9, Canada
[2] Univ Calgary, Seaman Family MR Res Ctr, Dept Radiol, Calgary, AB T2N 2T9, Canada
[3] Univ Calgary, Seaman Family MR Res Ctr, Dept Clin Neurosci, Calgary, AB T2N 2T9, Canada
关键词
fMRI; artifacts; filtering; Stockwell transform; Fourier transform; motion;
D O I
10.1002/mrm.10681
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A novel and automated technique is described for removing fMRI image artifacts resulting from motion outside of the imaging field of view. The technique is based on the Stockwell transform (ST), a mathematical operation that provides the frequency content at each time point within a time-varying signal. Using this technique, ID Fourier transforms (FTs) are performed on raw image data to obtain phase profiles. The time series of phase magnitude for each and every point in the phase profile is then subjected to the ST to obtain a time-frequency spectrum. The temporal location of an artifact is identified based on the magnitude of a frequency component relative to the median magnitude of that frequency's occurrence over all time points. After each artifact frequency is removed by replacing its magnitude with the median magnitude, an inverse ST is applied to regain the MR signal. Brain activity detection within fMRI datasets is improved by significantly reducing image artifacts that overlap anatomical regions of interest. The major advantage of ST-filtering is that artifact frequencies may be removed within a narrow time-window, while preserving the frequency information at all other time points. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:16 / 21
页数:6
相关论文
共 22 条
[1]  
Birn RM, 1999, HUM BRAIN MAPP, V7, P106, DOI 10.1002/(SICI)1097-0193(1999)7:2<106::AID-HBM4>3.0.CO
[2]  
2-O
[3]   Detection versus estimation in event-related fMRI: Choosing the optimal stimulus timing [J].
Birn, RM ;
Cox, RW ;
Bandettini, PA .
NEUROIMAGE, 2002, 15 (01) :252-264
[4]  
BRAMMER MJ, 2001, MAGNET RESON MED, V46, P344
[5]   IMPACT: Image-based physiological artifacts estimation and correction technique for functional MRI [J].
Chuang, KH ;
Chen, JH .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (02) :344-353
[6]  
Glover GH, 2000, MAGNET RESON MED, V44, P162, DOI 10.1002/1522-2594(200007)44:1<162::AID-MRM23>3.0.CO
[7]  
2-E
[8]  
GOODYEAR BG, 2000, P 10 ANN M ISMRM HON, P1419
[9]   REDUCTION OF SIGNAL FLUCTUATION IN FUNCTIONAL MRI USING NAVIGATOR ECHOES [J].
HU, XP ;
KIM, SG .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (05) :495-503
[10]   Comparing cortical activations for silent and overt speech using event-related fMRI [J].
Huang, J ;
Carr, TH ;
Cao, Y .
HUMAN BRAIN MAPPING, 2002, 15 (01) :39-53