LC determination of rofecoxib in bulk and pharmaceutical formulations

被引:21
作者
Radhakrishna, T [1 ]
Rao, DS [1 ]
Reddy, GO [1 ]
机构
[1] Dr Reddys Res Fdn, Dept Analyt Res, Hyderabad 500050, Andhra Pradesh, India
关键词
rofecoxib; related compounds; purity; assay; chlorophenyl methyl sulphone; formulation; RP-LC;
D O I
10.1016/S0731-7085(01)00493-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An isocratic reversed phase-liquid chromatographic (RP-LC) method has been developed for the determination and purity evaluation of rofecoxib in bulk and pharmaceutical dosage forms using photodiode array detection set at 225 nm. The method is simple, rapid and selective, The method is capable of detecting all process intermediates and other related compounds, which may be present at trace levels in finished products. Hence the method is very useful for process monitoring during the production of rofecoxib. Chlorophenyl methyl sulphone has been used as internal standard for the quantitative determination of rofecoxib. The method is linear in the range of 125-500 mug. The precision for inter- and intra-day assay variation of rofecoxib is below 1.6% relative standard deviation (R.S.D.). The accuracy determined as relative mean error (R.M.E.) for the intra-day assay is within +/-2.0%. The drug was extracted from tablets (Vioxx) using acetonitrile. The percentage recoveries from dosage forms were ranged from 98.2 to 102.6. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:617 / 628
页数:12
相关论文