Tubular-wire dual electrode for detection of thiols and disulfides by capillary electrophoresis/electrochemistry

被引:61
作者
Zhong, M
Lunte, SM
机构
[1] Univ Kansas, Dept Pharmaceut Chem, Lawrence, KS 66047 USA
[2] Univ Kansas, Dept Chem, Lawrence, KS 66047 USA
[3] Univ Kansas, Ctr Bioanalyt Res, Lawrence, KS 66047 USA
关键词
D O I
10.1021/ac9806198
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new dual-electrode detector for capillary electrophoresis is described. The detector consists of an integrated gold tubular electrode as the generator electrode and a gold wire electrode for detection. The detector configuration, including electrode size and position, has been optimized in terms of detection sensitivity and separation efficiency. After amalgamation of the dual electrode with mercury, the capillary electrophoresis/electrochemistry system was employed for simultaneous detection of thiols and disulfides. The response of cystine was found to be linear from 1 mu M to 1 mM with a LOD of 0.5 mu M (S/N = 3) and sensitivity of 60 pA/mu M. The detection limits represent 200-fold improvement over previously reported dual-electrode designs for the detection of disulfides. The use of this detector for identification of thiol- and disulfide-containing peptides was demonstrated with a tryptic digest of ribonuclease A.
引用
收藏
页码:251 / 255
页数:5
相关论文
共 27 条
[1]  
ALLISON LA, 1983, J CHROMATOGR, V255, P553
[2]   MICROCOLUMN SEPARATIONS OF SINGLE NERVE-CELL COMPONENTS [J].
EWING, AG .
JOURNAL OF NEUROSCIENCE METHODS, 1993, 48 (03) :215-224
[3]  
Ewing Andrew G., 1994, Analytical Chemistry, V66, P527
[4]  
Fermier AM, 1996, HRC-J HIGH RES CHROM, V19, P613, DOI 10.1002/jhrc.1240191104
[5]   Optimization of the separation and detection of the enantiomers of isoproterenol in microdialysis samples by cyclodextrin-modified capillary electrophoresis using electrochemical detection [J].
Hadwiger, ME ;
Torchia, SR ;
Park, S ;
Biggin, ME ;
Lunte, CE .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1996, 681 (02) :241-249
[6]  
Holland LA, 1998, ANAL COMMUN, V35, p1H
[7]   END-COLUMN DETECTION FOR CAPILLARY ZONE ELECTROPHORESIS [J].
HUANG, XH ;
ZARE, RN ;
SLOSS, S ;
EWING, AG .
ANALYTICAL CHEMISTRY, 1991, 63 (02) :189-192
[8]   DETERMINATION OF THIOLS AND DISULFIDES USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ELECTROCHEMICAL DETECTION [J].
KLEINMAN, WA ;
RICHIE, JP .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1995, 672 (01) :73-80
[9]   DUAL ELECTROCHEMICAL DETECTION OF CYSTEINE AND CYSTINE IN CAPILLARY ZONE ELECTROPHORESIS [J].
LIN, BL ;
COLON, LA ;
ZARE, RN .
JOURNAL OF CHROMATOGRAPHY A, 1994, 680 (01) :263-270
[10]   USE OF AN ON-COLUMN FRACTURE IN CAPILLARY ZONE ELECTROPHORESIS FOR SAMPLE INTRODUCTION [J].
LINHARES, MC ;
KISSINGER, PT .
ANALYTICAL CHEMISTRY, 1991, 63 (18) :2076-2078