Solution of the inverse problem of magnetic induction tomography (MIT)

被引:69
作者
Merwa, R
Hollaus, K
Brunner, P
Scharfetter, H
机构
[1] Graz Univ Technol, Inst Med Engn, A-8010 Graz, Austria
[2] Graz Univ Technol, Inst Fundamentals & Theory Elect Engn, A-8010 Graz, Austria
关键词
magnetic induction tomography; sensitivity; inverse problem; regularization; gradiometer;
D O I
10.1088/0967-3334/26/2/023
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Magnetic induction tomography (MIT) of biological tissue is used to reconstruct the changes in the complex conductivity distribution inside an object under investigation. The measurement principle is based on determining the perturbation Delta B of a primary alternating magnetic field B-0, which is coupled from an array of excitation coils to the object under investigation. The corresponding voltages Delta V and V-0 induced in a receiver coil carry the information about the passive electrical properties (i.e. conductivity, permittivity and permeability). The reconstruction of the conductivity distribution requires the solution of a 3D inverse eddy current problem. As in EIT the inverse problem is ill-posed and on this account some regularization scheme has to be applied. We developed an inverse solver based on the GaussNewton-one-step method for differential imaging, and we implemented and tested four different regularization schemes: the first and second approaches employ a classical smoothness criterion using the unit matrix and a differential matrix of first order as the regularization matrix. The third method is based on variance uniformization, and the fourth method is based on the truncated singular value decomposition. Reconstructions were carried out with synthetic measurement data generated with a spherical perturbation at different locations within a conducting cylinder. Data were generated on a different mesh and 1% random noise was added. The model contained 16 excitation coils and 32 receiver coils which could be combined pairwise to give 16 planar gradiometers. With 32 receiver coils all regularization methods yield fairly good 3D-images of the modelled changes of the conductivity distribution, and prove the feasibility of difference imaging with MIT. The reconstructed perturbations appear at the right location, and their size is in the expected range. With 16 planar gradiometers an additional spurious feature appears mirrored with respect to the median plane with negative sign. This demonstrates that a symmetrical arrangement with one ring of planar gradiometers cannot distinguish between a positive conductivity change at the true location and a negative conductivity change at the mirrored location.
引用
收藏
页码:S241 / S250
页数:10
相关论文
共 23 条
[1]  
Brandstätter B, 2003, PHYSIOL MEAS, V24, P437, DOI 10.1088/0967-3334/24/355
[2]   Measurement of liver iron overload by magnetic induction using a planar gradiometer:: preliminary human results [J].
Casañas, R ;
Scharfetter, H ;
Altes, A ;
Remacha, A ;
Sarda, P ;
Sierra, J ;
Merwa, R ;
Hollaus, K ;
Rosell, J .
PHYSIOLOGICAL MEASUREMENT, 2004, 25 (01) :315-323
[3]  
Cheney M, 1990, Int J Imaging Syst Technol, V2, P66, DOI 10.1002/ima.1850020203
[4]   Regularized reconstruction in electrical impedance tomography using a variance uniformization constraint [J].
CohenBacrie, C ;
Goussard, Y ;
Guardo, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (05) :562-571
[6]   Magnetic induction tomography [J].
Griffiths, H .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2001, 12 (08) :1126-1131
[7]   Magnetic induction tomography - A measuring system for biological tissues [J].
Griffiths, H ;
Stewart, WR ;
Gough, W .
ELECTRICAL BIOIMPEDANCE METHODS: APPLICATIONS TO MEDICINE AND BIOTECHNOLOGY, 1999, 873 :335-345
[8]  
Hanke M., 1993, Surveys on Mathematics for Industry, V3, P253
[9]   Fast calculation of the sensitivity matrix in magnetic induction tomography by tetrahedral edge finite elements and the reciprocity theorem [J].
Hollaus, K ;
Magele, C ;
Merwa, R ;
Scharfetter, H .
PHYSIOLOGICAL MEASUREMENT, 2004, 25 (01) :159-168
[10]  
HOLLAUS K, 2002, P 10 IGTE S, V25, P159