Application of high TcSQUID magnetometer for sentinel-lymph node biopsy

被引:18
作者
Tanaka, S [1 ]
Hirata, A
Saito, Y
Mizoguchi, T
Tamaki, Y
Sakita, I
Monden, M
机构
[1] Toyohashi Univ Technol, Toyohashi, Aichi 4418580, Japan
[2] Osaka Univ, Grad Sch Med, Dept Surg & Clin Oncol, Suita, Osaka 5650871, Japan
关键词
SQUID; biopsy; detection; lymph node;
D O I
10.1109/77.919432
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The basic performance of a nanoparticle detection system for lymph node used with a high-Tc superconducting quantum interference device (SQUID) was investigated. Ultra-small iron oxide particles of 360 pg could be detected with spacing of 1 mm between the SQUID magnetometer and the particles, When the space was widened to 40 mm, the detectable weight of the particles was increased and was 1.6 mug.
引用
收藏
页码:665 / 668
页数:4
相关论文
共 13 条
[1]  
[Anonymous], ANN SURG
[2]   Guidelines for sentinel node biopsy and lymphatic mapping of patients with breast cancer [J].
Cox, CE ;
Pendas, S ;
Cox, JM ;
Joseph, E ;
Shons, AR ;
Yeatman, T ;
Ku, NN ;
Lyman, GH ;
Berman, C ;
Haddad, F ;
Reintgen, DS .
ANNALS OF SURGERY, 1998, 227 (05) :645-653
[3]  
ENGELHARDT R, 1996, P 10 INT C BIOM
[4]   Detection of magnetic nanoparticles with superconducting quantum interference device (SQUID) magnetometer and application to immunoassays [J].
Enpuku, K ;
Minotani, T ;
Gima, T ;
Kuroki, Y ;
Itoh, Y ;
Yamashita, M ;
Katakura, Y ;
Kuhara, S .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS, 1999, 38 (10A) :L1102-L1105
[5]  
HASEGAWA M, 1992, P 6 INT C FERR KYOT, P1007
[6]  
HOROWITZ P, 1995, ART ELECT, P1032
[7]  
KOETITZ R, 1997, IEEE T APPL SUPERCON, V7, P3678
[8]   Nuclear magnetic resonance using a high temperature superconducting quantum interference device [J].
Kumar, S ;
Matthews, R ;
Haupt, SG ;
Lathrop, DK ;
Takigawa, M ;
Rozen, JR ;
Brown, SL ;
Koch, RH .
APPLIED PHYSICS LETTERS, 1997, 70 (08) :1037-1039
[9]   A SQUID based AC susceptometer for the investigation of large samples [J].
Pasquarelli, A ;
DelGratta, C ;
DellaPenna, S ;
DiLuzio, S ;
Pizzella, V ;
Romani, GL .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2533-2539
[10]   Low-field magnetic resonance imaging with a high-Tc dc superconducting quantum interference device [J].
Schlenga, K ;
McDermott, R ;
Clarke, J ;
de Souza, RE ;
Wong-Foy, A ;
Pines, A .
APPLIED PHYSICS LETTERS, 1999, 75 (23) :3695-3697