Analysis of trace elements in the central nerve tissues with inductively coupled plasma-mass spectrometry

被引:11
作者
Ejima, A
Watanabe, C
Koyama, H
Satoh, H
机构
[1] Dept. of Environ. Health Sciences, Tohoku University, School of Medicine
[2] Dept. of Environ. Health Sciences, Tohoku Univerty, School of Medicine, Aoba-ku, Sendai 980-77
关键词
inductively coupled plasma-mass spectrometry; spectral interferences; nonspectral interferences; spinal cord; amyotrophic lateral sclerosis; AMYOTROPHIC-LATERAL-SCLEROSIS; SPINAL-CORDS; MANGANESE; SELENIUM;
D O I
10.1620/tjem.178.1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To investigate the roles of various elements in neurological disorders, multi-element analysis of limited quantities of samples in the central nervous system is required. Inductively coupled plasma-mass spectrometry (ICP-MS) analysis suits this requirement, but spectral and non-spectral interferences are inevitable. We studied correction methods for the non-spectral interferences by analyzing signals of 21 elements in various concentrations of HNO3 as well as 5 major elements (Na, K, P, Ca, and Cl). Using internal standards, the interferences caused by the major elements were corrected, but the interferences caused by HNO3 were impossible to correct for the elements with high ionization potentials. Thus, me decided to use a standard addition method to correct these interferences. The spectral interferences on Mn and Fe rising from HNO3 or major elements were compared with the signals of samples. Although the interferences on Mn mere negligible, those on Pe were considerable and careful blank subtraction were needed. We measured concentrations of elements in the spinal cord of 4 controls and a patient with amyotrophic lateral sclerosis (ALS) using the standard addition method. The concentrations of Mn, Se, Fe, and Zn in the controls mere nearly the same as previously reported values, whereas Mn concentrations in the ALS patient mere higher.
引用
收藏
页码:1 / 10
页数:10
相关论文
共 14 条
[1]   DETERMINATION OF TRACE-METALS IN MARINE BIOLOGICAL REFERENCE MATERIALS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
WILLIE, SN ;
BERMAN, SS .
ANALYTICAL CHEMISTRY, 1988, 60 (07) :687-691
[2]  
EJIMA A, 1995, BIOMED RES TRACE ELE, V16, P179
[3]   INTERFERENCE EFFECTS AND THEIR CONTROL IN ICP-MS ANALYSIS OF SERUM AND SALINE SOLUTIONS [J].
GEROTTO, M ;
DELLANDREA, E ;
BORTOLI, A ;
MARCHIORI, M ;
PALONTA, M ;
TRONCON, A .
MICROCHEMICAL JOURNAL, 1995, 51 (1-2) :73-87
[4]   A MATHEMATICAL CORRECTION METHOD FOR SPECTRAL INTERFERENCES ON SELENIUM IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
GOOSSENS, J ;
MOENS, L ;
DAMS, R .
TALANTA, 1994, 41 (02) :187-193
[5]   ELIMINATION OF INTERFERENCES IN THE DETERMINATION OF ARSENIC AND SELENIUM IN BIOLOGICAL SAMPLES BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
GOOSSENS, J ;
VANHAECKE, F ;
MOENS, L ;
DAMS, R .
ANALYTICA CHIMICA ACTA, 1993, 280 (01) :137-143
[6]  
KHARE SS, 1990, NEUROTOXICOLOGY, V11, P521
[7]   DETERMINATION OF MANGANESE CONCENTRATIONS IN THE SPINAL-CORDS FROM AMYOTROPHIC-LATERAL-SCLEROSIS PATIENTS BY INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROSCOPY [J].
KIHIRA, T ;
MUKOYAMA, M ;
ANDO, K ;
YASE, Y ;
YASUI, M .
JOURNAL OF THE NEUROLOGICAL SCIENCES, 1990, 98 (2-3) :251-258
[8]   MANGANESE, SELENIUM AND OTHER TRACE-ELEMENTS IN SPINAL-CORD, LIVER AND BONE IN MOTOR-NEURON DISEASE [J].
MITCHELL, JD ;
EAST, BW ;
HARRIS, IA ;
PENTLAND, B .
EUROPEAN NEUROLOGY, 1991, 31 (01) :7-11
[9]   INCREASED MANGANESE LEVEL IN SPINAL CORDS OF AMYOTROPHIC LATERAL SCLEROSIS DETERMINED BY RADIOCHEMICAL NEUTRON-ACTIVATION ANALYSIS [J].
MIYATA, S ;
NAKAMURA, S ;
NAGATA, H ;
KAMEYAMA, M .
JOURNAL OF THE NEUROLOGICAL SCIENCES, 1983, 61 (02) :283-293
[10]   Microwave digestion using dual PTFE containers for analysis of trace elements in small amounts of biological samples [J].
Mizushima, R ;
Yonezawa, M ;
Ejima, A ;
Koyama, H ;
Satoh, H .
TOHOKU JOURNAL OF EXPERIMENTAL MEDICINE, 1996, 178 (01) :75-79