Intraoral radiographic storage phosphor image mean pixel values and signal-to-noise ratio - Effects of calibration

被引:25
作者
Hayakawa, Y
Farman, AG
Kelly, MS
Kuroyanagi, K
机构
[1] Tokyo Dent Coll, Dept Oral & Maxillofacial Radiol, Mihama Ku, Chiba 2618502, Japan
[2] Univ Louisville, Sch Dent, Div Radiat & Imaging Sci, Louisville, KY 40292 USA
[3] Univ Louisville, Hlth Sci Ctr, Radiat Safety Off, Louisville, KY 40292 USA
来源
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS | 1998年 / 86卷 / 05期
关键词
D O I
10.1016/S1079-2104(98)90354-7
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. The DIGORA intraoral radiographic storage phosphor system needs to be calibrated before images are made. Calibration involves inputting of the maximum exposure to be used. This investigation studied the effects of different maximum exposure calibration settings on the mean pixel value for selected regions of interest and the signal-to-noise ratio for images of a test phantom. Study design. A dental phantom containing a step wedge made of different thicknesses of homogeneously radiopaque bone-equivalent material was imaged at 70 kVp with exposures ranging from 12.8 to 105.2 mu C . kg(-1). Images were displayed through use of imaging software, and a region of interest was set for each bone-step. The mean pixel values and their standard deviations were measured. Results. Except for very low exposures (< 10% of the calibrated maximum), there was a linear relationship between exposure and the pixel values within the regions of interest irrespective of the calibrated maximum exposure. Low exposures resulted in underexposed low-contrast images. Low calibrated maximum exposures (< 54.5.mu C kg(-1)) resulted in low-density images with poor signal-to-noise ratios. Conclusions. Because of the very wide image latitude of the DIGORA system, loss of image quality was not observed as a result of adjustments in the calibration setting over the range of exposures commonly used in dental practice. The highest accepted exposure was limited by prior calibration; hence, if diligence is applied, patient exposure can be minimized without detriment to image quality.
引用
收藏
页码:601 / 605
页数:5
相关论文
共 11 条
[1]   On the dynamic range of different X-ray photon detectors in intra-oral radiography. A comparison of image quality in film, charge-coupled device and storage phosphor systems [J].
Borg, E ;
Grondahl, HG .
DENTOMAXILLOFACIAL RADIOLOGY, 1996, 25 (02) :82-88
[2]  
Campbell RC, 1967, STAT BIOL
[3]  
Kapa S F, 1990, Dentomaxillofac Radiol, V19, P67
[4]   INTRAORAL COMPUTED RADIOGRAPHY USING THE FUJI COMPUTED RADIOGRAPHY IMAGING PLATE - CORRELATION BETWEEN IMAGE QUALITY AND READING CONDITION [J].
KASHIMA, I ;
SAKURAI, T ;
MATSUKI, T ;
NAKAMURA, K ;
AOKI, H ;
ISHII, M .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1994, 78 (02) :239-246
[5]   COMPUTED PANORAMIC TOMOGRAPHY WITH SCANNING LASER-STIMULATED LUMINESCENCE [J].
KASHIMA, I ;
KANNO, M ;
HIGASHI, T ;
TAKANO, M .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1985, 60 (04) :448-453
[6]   COMPUTED RADIOGRAPHY WITH PHOTOSTIMULABLE PHOSPHOR IN ORAL AND MAXILLOFACIAL RADIOLOGY [J].
KASHIMA, I .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1995, 80 (05) :577-598
[7]  
MATSUDA Y, 1995, ORAL RADIOL, V11, P11
[8]  
*OR SOR, 1995, DIGORA OP INSTR
[9]   COMPUTED RADIOGRAPHY UTILIZING SCANNING LASER STIMULATED LUMINESCENCE [J].
SONODA, M ;
TAKANO, M ;
MIYAHARA, J ;
KATO, H .
RADIOLOGY, 1983, 148 (03) :833-838
[10]   BASIC TECHNICAL PROPERTIES OF A SYSTEM FOR DIRECT ACQUISITION OF DIGITAL INTRAORAL RADIOGRAPHS [J].
WELANDER, U ;
NELVIG, P ;
TRONJE, G ;
MCDAVID, WD ;
DOVE, SB ;
MORNER, AC ;
CEDERLUND, T .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 1993, 75 (04) :506-516