Interference by cellular telephone with an implantable loop recorder

被引:2
作者
Trigano, A [1 ]
Blandeau, O [1 ]
Levy, S [1 ]
机构
[1] Ctr Hosp Univ Nord, Dept Cardiol, F-13915 Marseille, France
关键词
cellular phone interference; electromagnetic interference; implantable cardiac device; implantable loop recorder;
D O I
10.1007/s10840-005-0303-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report the observation of interference between an implanted loop recorder and a cellular telephone when the latter was placed over the subcutaneous pocket. The artifact produced during the ringing of the telephone was automatically recorded and stored in the device memory. A similar interference was documented during ex vivo testing.
引用
收藏
页码:237 / 240
页数:4
相关论文
共 7 条
[1]   Electromagnetic interference of an implantable loop recorder by commonly encountered electronic devices [J].
de Cock, CC ;
Spruijt, HJ ;
van Campen, LMC ;
Plu, AW ;
Visser, CA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (10) :1516-1518
[2]  
Gimbel J R, 2003, Heart, V89, pe10, DOI 10.1136/heart.89.3.e10
[3]   Interference with cardiac pacemakers by cellular telephones [J].
Hayes, DL ;
Wang, PJ ;
Reynolds, DW ;
Estes, NAM ;
Griffith, JL ;
Steffens, RA ;
Carlo, GL ;
Findlay, GK ;
Johnson, CM .
NEW ENGLAND JOURNAL OF MEDICINE, 1997, 336 (21) :1473-1479
[4]   Electromagnetic interference of pacemakers by mobile phones [J].
Irnich, W ;
Batz, L ;
Muller, R ;
Tobisch, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1431-1446
[5]  
*MEDTR INC, 2001, TECHN MAN REV PLUS 9, V2
[6]   Intermittent pacemaker dysfunction caused by digital mobile telephones [J].
Naegeli, B ;
Osswald, S ;
Deola, M ;
Burkart, F .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 27 (06) :1471-1477
[7]   Arrhythmia detection by patient and auto-activation in implantable loop recorders [J].
Ng, E ;
Stafford, PJ ;
Ng, GA .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2004, 10 (02) :147-152