Quantitative α-particle detection in a homogeneous medium with LR115

被引:10
作者
Pape, A [1 ]
Adloff, JC
Barillon, R
Haessler, A
Hoernel, A
Nourreddine, A
Oster, D
Weidmann, D
机构
[1] Inst Rech Subatom, F-67037 Strasbourg 2, France
[2] Fac Sci, Lab Phys & Tech Nucl, El Jadida, Morocco
关键词
LR115; detection efficiency; alpha-dosimetry; radon dosimetry;
D O I
10.1016/S0168-583X(98)00398-X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Agreement between theoretical and experimental track densities in LR115 cellulose nitrate detectors shows that alpha-particles can be detected quantitatively in air, solutions, and finely divided solids. For general use, a readily applicable form of the numerically integrated track registration equation for LR115 contact autoradiography is presented. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:557 / 560
页数:4
相关论文
共 7 条
  • [1] Track etch velocity study in a radon detector (LR 115, cellulose nitrate)
    Barillon, R
    Fromm, M
    Chambaudet, A
    Marah, H
    Sabir, A
    [J]. RADIATION MEASUREMENTS, 1997, 28 (1-6) : 619 - 628
  • [2] BENARD J, 1958, NOUVEAU TRAITE CHIMI, V4, P424
  • [3] BIERSACK JP, 1991, CODE TRIM 91
  • [4] KRETZ R, 1978, CRC HDB CHEM PHYSICS, pB223
  • [5] MILES JCH, 1986, NUCL TRACKS RAD MEAS, V12, P735
  • [6] MULLER GO, 1965, PRAKTIKUM QUANTITATI, P257
  • [7] SINGH M, 1986, NUCL TRACKS RAD MEAS, V12, P739