Enzymatic assays for creatinine: Time for action

被引:40
作者
Panteghini, Mauro [1 ,2 ]
机构
[1] Univ Milan, Ctr Metrol Traceabil Lab Med CIRME, Milan, Italy
[2] Univ Milan, Dept Clin Sci Luigi Sacco, Milan, Italy
关键词
calibration; glomerular filtration rate; isotope dilution-mass spectrometry; kidney function tests; reference standards; traceability;
D O I
10.1080/00365510802149978
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Estimation of glomerular filtration rate (eGFR) on the basis of serum creatinine concentration measurements using equations is critical to ongoing global public health efforts to improve the diagnosis and treatment of chronic kidney disease. There is now ongoing activity to promote world-wide standardization of methods to measure creatinine concentrations, together with the introduction of a revised eGFR equation appropriate for use with standardized creatinine methods. Standardization of calibration, i.e. implementation of calibration traceable to higher-order reference measurement procedures and reference materials, does not, however, correct for analytical interferences of field methods (non-specificity bias). To account for the sensitivity of alkaline picrate-based methods to non-creatinine chromogens, some manufacturers have adjusted the calibration to minimize the pseudo-creatinine contribution of plasma proteins, thereby producing results more closely aligned with the reference method (isotope dilution-mass spectrometry), but this strategy makes the assumption that the non-creatinine chromogen interference is constant among samples, which is an oversimplification. Thus, analytical non-specificity for substances found in individual patient samples affects the accuracy of eGFR computed from serum creatinine concentrations for any alkaline picrate method, including the so-called "compensated'' Jaffe methods. Using assays that are more specific for serum creatinine, such as those based on enzymatic reactions, may provide more reliable eGFR values. Supporting the choice of more specific assays by clinical laboratories is one of the main tasks of our profession in achieving the ultimate clinical goal, which is to routinely report an accurate eGFR in all pertinent clinical situations.
引用
收藏
页码:84 / 88
页数:5
相关论文
共 28 条
[1]  
[Anonymous], RIMEL IJLAM S
[2]  
[Anonymous], CLIN CHEM LAB MED
[3]  
BLIJENBERG BG, 1995, EUR J CLIN CHEM CLIN, V33, P855
[4]   Reference intervals for serum creatinine concentrations: Assessment of available data for global application [J].
Ceriotti, Ferruccio ;
Boyd, James C. ;
Klein, Gerhard ;
Henny, Joseph ;
Queralto, Josep ;
Kairisto, Veli ;
Panteghini, Mauro .
CLINICAL CHEMISTRY, 2008, 54 (03) :559-566
[5]  
Ceriotti Ferruccio, 2007, Clin Biochem Rev, V28, P115
[6]   Quality of serum creatinine measurement in light of EQA programs [J].
Friedecky, Bedrich ;
Kratochvila, Josef ;
Budina, Marek ;
Jabor, Antonin .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2007, 45 (05) :685-688
[7]  
Infusino I, 2007, CLIN CHEM LAB MED, V45, pS275
[8]  
Infusino I, 2007, CLIN CHEM LAB MED, V45, pA155
[9]  
ISO, ISO 17511: 2003
[10]   Determination of reference intervals for serum creatinine, creatinine excretion and creatinine clearance with an enzymatic and a modified Jaffe method [J].
Junge, W ;
Wilke, B ;
Halabi, A ;
Klein, G .
CLINICA CHIMICA ACTA, 2004, 344 (1-2) :137-148