A real-time gyroscopic system for three-dimensional measurement of lumbar spine motion

被引:83
作者
Lee, RYW
Laprade, J
Fung, EHK
机构
[1] Hong Kong Polytech Univ, Dept Rehabil Sci, Hunghom, Hong Kong, Peoples R China
[2] Hong Kong Polytech Univ, Dept Mech Engn, Hunghom, Hong Kong, Peoples R China
关键词
kinematics; lumbar spine; inertial tracking device; gyroscopes;
D O I
10.1016/S1350-4533(03)00115-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
This paper describes a new method of measuring the three-dimensional movements of the lumbar spine in real time. The measurement system consisted of solid-state gyroscopes which were attached to the trunk. They measured the angular rates of rotations in three dimensions, which were then integrated to obtain the orientation. The sensors contained gravitometers and magnetometers which provided additional information for eliminating any drift of the gyroscopes. The reliability of the data provided by the gyroscopic system was examined in a group of 19 young healthy subjects. The similarity of the movement-time curves obtained in three repeated measurements was assessed by the coefficient of multiple correlation. The coefficients were found to be high, ranging from 0.972 to 0.991. The reliability of the data was slightly lower for measuring axial rotation. The device did not only quantify the kinematic patterns in the primary plane of movements, but also the accompanying movements in the other planes. Flexion and extension was found to be mainly confined to the sagittal plane, whereas lateral bending and axial rotation always accompanied each other. It was concluded that the inertial tracking device would be a useful tool for clinical measurement as well as biomechanical investigations. (C) 2003 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:817 / 824
页数:8
相关论文
共 33 条
[1]
HAS THE LUMBAR SPINE A MARGIN OF SAFETY IN FORWARD BENDING [J].
ADAMS, MA ;
HUTTON, WC .
CLINICAL BIOMECHANICS, 1986, 1 (01) :3-6
[2]
APPLICATION OF A MAGNETIC TRACKING DEVICE TO KINESIOLOGIC STUDIES [J].
AN, KN ;
JACOBSEN, MC ;
BERGLUND, LJ ;
CHAO, EYS .
JOURNAL OF BIOMECHANICS, 1988, 21 (07) :613-&
[4]
SPINAL ANALYSIS USING A 3-DIMENSIONAL RADIOGRAPHIC TECHNIQUE [J].
BROWN, RH ;
BURSTEIN, AH ;
NASH, CL ;
SCHOCK, CC .
JOURNAL OF BIOMECHANICS, 1976, 9 (06) :355-365
[5]
Accuracy of an electromagnetic measurement device and application to the measurement and description of knee joint motion [J].
Bull, AMJ ;
Berkshire, FH ;
Amis, AA .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1998, 212 (H5) :347-355
[6]
Three-dimensional measurement of lumbar spine kinematics for fast bowlers in cricket [J].
Burnett, AF ;
Barrett, CJ ;
Marshall, RN ;
Elliott, BC ;
Day, RE .
CLINICAL BIOMECHANICS, 1998, 13 (08) :574-583
[7]
DVORAK J, 1991, SPINE, V16, P562
[8]
A JOINT COORDINATE SYSTEM FOR THE CLINICAL DESCRIPTION OF 3-DIMENSIONAL MOTIONS - APPLICATION TO THE KNEE [J].
GROOD, ES ;
SUNTAY, WJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1983, 105 (02) :136-144
[9]
Repeated measures of adult normal walking using a video tracking system [J].
Growney, E ;
Meglan, D ;
Johnson, M ;
Cahalan, T ;
An, KN .
GAIT & POSTURE, 1997, 6 (02) :147-162
[10]
Helliwell P, 1992, LUMBAR SPINE BACK PA, P173